4–7
Rapid fFN 10Q
System User Manual
Section 4 —
Software Functions — TEST PATIENT
The analyzer automatically compares the CASSETTE LOT# used to set
calibration with the cassette lot number used for patient testing. If the
LOT NUMBERS do not match, the test process cannot continue. When
this occurs, the cassette lot number used for calibration will be displayed,
and the user is prompted to recalibrate the system.
Enter up to 16 alphanumeric characters for a PATIENT ID and press
ENTER. The PATIENT ID must be entered to proceed to the next step.
This message will be displayed if a cassette is present in the analyzer
prior to reaching the next screen. Remove cassette and press ENTER.
CASSETTE LOT CHANGED
CALIBRATE SYSTEM
L1002
ESC-MAIN MENU
PATIENT ID
>JANE DOE 123
ENTER-ACCEPT
REMOVE CASSETTE
PRESS ENTER
TO CONTINUE