Chapter 4: Commissioning
4. If the calibration was not successful, “Error”
will appear.
In the event of an error, check the entire sys-
tem. Resistance in the overall system may be
too high. Then rerun the calibration.
ST
Tuesday l 22.03.2017 l 20:36
40
30
20
10
0
Calibrate Tube
100 l/min
Error
Press the MFK to continue
If you do not perform a tube calibration, the calibration data from the last tube calibra-
tion are used. Therapy can be started as usual, but the pressure measurement may be
incorrect. We recommend always performing tube calibration.
When commissioning the device for the first time without calibrating it, the default cali-
bration data
1
stored in the device are used.
1
Operation without humidifier:
Leakage tube circuit:
∅
22 mm; 1.80 m
Height: ca. 44 m NHN, air pressure: approx. 1008 hPa
Connected accessories: no
Operation with humidifier:
Leakage tube circuit:
∅
22 mm; 1.80 m
Height: ca. 44 m NHN, air pressure: approx. 1008 hPa
Connected accessories: Humidifier AquaTREND III-NG
Содержание 5CPD10
Страница 2: ......
Страница 10: ...10 List of figures ...
Страница 11: ...Chapter 1 Introduction 11 Chapter 1 Introduction ...
Страница 21: ...Chapter 2 Safety Information 21 Chapter 2 Safety Information ...
Страница 28: ...28 Chapter 2 Safety Information ...
Страница 29: ...Chapter 3 Description of Device 29 Chapter 3 Description of Device ...
Страница 36: ...36 Chapter 3 Description of Device ...
Страница 37: ...Chapter 4 Commissioning 37 Chapter 4 Commissioning ...
Страница 56: ...56 Chapter 4 Commissioning ...
Страница 57: ...Chapter 5 Therapy Modes 57 Chapter 5 Therapy Modes ...
Страница 65: ...Chapter 6 Operating the Device 65 Chapter 6 Operating the Device ...
Страница 91: ...Chapter 7 Notifications Messages 91 Chapter 7 Notifications Messages ...
Страница 95: ...Chapter 8 Cleaning and Disinfection 95 Chapter 8 Cleaning and Disinfection ...
Страница 104: ...104 Chapter 8 Cleaning and Disinfection ...
Страница 105: ...Chapter 9 Annex 105 Chapter 9 Annex ...