Chapter 8: Cleaning and Disinfection
103
Before using the device for another patient you must complete the following procedures:
Component
What action is required?
Bacterial filter
Replace
Coarse and fine filter
Replace
Mask
Replace or recondition according to the manufacturer’s
guidelines if possible
Tube circuit
Replace or recondition according to the manufacturer’s guide-
lines if possible
Humidifier
Recondition according to the manufacturer’s guidelines
Device
Disinfect the device surfaces with a disposable germicidal
wipe.
We recommend an alcohol-free surface disinfectant,
e.g. schülke mikrozid sensitive wipes
. Make sure that no liquid
is drawn into the device through the openings.
Let the disinfectant soak in according to the disinfectant man-
ufacturer’s instructions.
Wipe any disinfectant agent residues from the device using
a disposable wipe.
Содержание 5CPD10
Страница 2: ......
Страница 10: ...10 List of figures ...
Страница 11: ...Chapter 1 Introduction 11 Chapter 1 Introduction ...
Страница 21: ...Chapter 2 Safety Information 21 Chapter 2 Safety Information ...
Страница 28: ...28 Chapter 2 Safety Information ...
Страница 29: ...Chapter 3 Description of Device 29 Chapter 3 Description of Device ...
Страница 36: ...36 Chapter 3 Description of Device ...
Страница 37: ...Chapter 4 Commissioning 37 Chapter 4 Commissioning ...
Страница 56: ...56 Chapter 4 Commissioning ...
Страница 57: ...Chapter 5 Therapy Modes 57 Chapter 5 Therapy Modes ...
Страница 65: ...Chapter 6 Operating the Device 65 Chapter 6 Operating the Device ...
Страница 91: ...Chapter 7 Notifications Messages 91 Chapter 7 Notifications Messages ...
Страница 95: ...Chapter 8 Cleaning and Disinfection 95 Chapter 8 Cleaning and Disinfection ...
Страница 104: ...104 Chapter 8 Cleaning and Disinfection ...
Страница 105: ...Chapter 9 Annex 105 Chapter 9 Annex ...