Chapter 1: Introduction
The TREND III respiratory therapy device is available in the following types: CPAP, BILEVEL
S, BILEVEL ST, AutoCPAP, AutoBILEVEL S and AutoBILEVEL ST.
All TREND III devices have a respiratory trigger. The respiratory trigger recognises the
patient’s difficulties in inhaling or exhaling and notifies the control equipment. A vol-
ume-based trigger can be set for inhaling, while a flow-based trigger is available for
exhaling (Values not adjustable for CPAP and AutoCPAP devices). In combination with
optimal slope adjustment, the device ensures very comfortable therapy for patients, who
feel like the device is gently supporting their natural breathing.
The following therapy modes are available with the TREND III, depending on the device
type.
Overview of the device types and modes
Typ / Modus
CPAP
APAP
S
T
ST
A-S
TREND III CPAP
•
TREND III AutoCPAP
•
•
TREND III BILEVEL S
•
•
TREND III BILEVEL ST
•
•
•
•
TREND III AutoBILEVEL S
•
•
•
•
TREND III AutoBILEVEL ST
•
•
•
•
•
•
Description therapy modes
Mode
CPAP
For therapy, the device provides continuous positive pressure.
APAP
APAP mode is an extension to CPAP mode and offers the additional function of
automatic adjustment to the therapeutic pressure (CPAP) within specific limits.
In addition, the speed of the pressure increase can be adapted to the patient’s
special needs in 5 stages.
S
The
transition from one pressure level to the next is triggered when the patient
begins to breathe. The pressure is decreased to the set lower level when the
inspiration flow ends. In addition, a safety frequency (S frequency) can be acti-
vated. This ensures patients are still ventilated with the set parameters in case
of respiratory arrest.
A-S
Auto S mode is an extension to S mode and offers the additional function of
automatic adjustment to the expiratory pressure (EPAP) within specific limits.
The IPAP is also increased so that the distance between the pressure levels
(pressure difference between IPAP and EPAP) always remains the same. In
addition, the speed of the pressure increase can be adapted to the patient’s
special needs in 5 stages.
Содержание 5CPD10
Страница 2: ......
Страница 10: ...10 List of figures ...
Страница 11: ...Chapter 1 Introduction 11 Chapter 1 Introduction ...
Страница 21: ...Chapter 2 Safety Information 21 Chapter 2 Safety Information ...
Страница 28: ...28 Chapter 2 Safety Information ...
Страница 29: ...Chapter 3 Description of Device 29 Chapter 3 Description of Device ...
Страница 36: ...36 Chapter 3 Description of Device ...
Страница 37: ...Chapter 4 Commissioning 37 Chapter 4 Commissioning ...
Страница 56: ...56 Chapter 4 Commissioning ...
Страница 57: ...Chapter 5 Therapy Modes 57 Chapter 5 Therapy Modes ...
Страница 65: ...Chapter 6 Operating the Device 65 Chapter 6 Operating the Device ...
Страница 91: ...Chapter 7 Notifications Messages 91 Chapter 7 Notifications Messages ...
Страница 95: ...Chapter 8 Cleaning and Disinfection 95 Chapter 8 Cleaning and Disinfection ...
Страница 104: ...104 Chapter 8 Cleaning and Disinfection ...
Страница 105: ...Chapter 9 Annex 105 Chapter 9 Annex ...