-37-
MN1-1014 Rev. 14
5-3. Sterilization
Perform following sterilization as necessary.
Probe
Puncture adapter
(Needle guide)
Puncture adapter
(Knob)
Needle stopper Depth gauge
EOG sterilization
○
○
○
○
○
Liquid sterilization
○
○
Not compatible Not compatible Not compatible
5-3-1. Ethylene oxide gas (EOG) sterilization
Applicable gases
General name
Trade name
Manufacturer
Ethylene oxide gas
AMPROLENE
TM
84% density
Central Uni Co., LTD.
Sterilization procedure
Washing
Be sure to wash and fully dry the equipment before sterilization.
See section 5-1.
EOG sterilization
Sterilize the following conditions:
Temperature: 10
℃
to 60
℃
50°F to 140°F
Humidity: 30
%
to 85
%
Compression
to +101.175kPa
Decompression
to -100.000kPa
Aeration
Aerate the following conditions to ensure that no gases remain.
Temperature:
10°C to 60°C
50°F to 140°F
Humidity:
30
%
to 85
%
Compression
to +101.175kPa
Decompression
to -100.000kPa
Period:
5 days (natural aeration)
Warning
Perform full aeration after gas sterilization.
Residual gas can cause an adverse reaction on the bodies of the operator or patient.