-35-
MN1-1014 Rev. 14
5-2. Disinfection
Perform following disinfection as necessary.
Probe
Puncture adapter
(Needle guide)
Puncture adapter
(Knob)
Needle stopper Depth gauge
Liquid disinfection
○
Perform sterilization the puncture adapter or other equipment.
Gas disinfection
○
5-2-1. Chemical disinfection
Applicable chemicals
General name
Trade name
Manufacturer
Glutaral
CIDEX™
Solution 2.4%
ADVANCED STERILIZATION PRODUCTS
®
A Johnson & Johnson company
Division of Ethicon, Inc.
Ortho-phthalaldehyde
CIDEX OPA™
Solution 0.55%
Glutaral
STERIHYDE™
Practical liquid 2W/V%
Maruishi Pharmaceutical Co., Ltd.
Benzalkonium chloride
DETERGICIDE™
Practical liquid 0.2W/V%
Yufu Itonaga Co., Ltd.
Benzethonium chloride
HYAMINE™
Practical liquid 0.1W/V%
DAIICHI SANKYO Co., Ltd.
Disinfection procedure
Washing
Be sure to wash and fully dry the probe before disinfection.
See section 5-1.
Soaking in
chemical solution
Soak the probe tip in a chemical solution at the shortest time where
effective disinfection can be achieved.
Water rinsing
Sufficiently rinse the probe using sterilized water to remove chemicals.
Remarks
Soaking the probe tip in CIDEX OPA™ solution 0.55% may result in discoloration of the silicone, but this
does not affect performance or safety.
Warning
After chemical disinfection, thoroughly wash the probe with sterilized water.
Residual chemicals can cause an adverse reaction on the bodies of the operator or patient.
○: Compatible methods
Содержание UST-9113P-3.5
Страница 2: ...MN1 1014 Rev 14 ii ...
Страница 16: ... 10 MN1 1014 Rev 14 2 Storage case Label A Label B ...
Страница 18: ... 12 MN1 1014 Rev 14 ...
Страница 24: ... 18 MN1 1014 Rev 14 ...
Страница 46: ... 40 MN1 1014 Rev 14 ...
Страница 48: ... 42 MN1 1014 Rev 14 ...
Страница 52: ... 46 MN1 1014 Rev 14 ...
Страница 54: ... 48 MN1 1014 Rev 14 ...