-27-
MN1-1014 Rev. 14
4-5. Removal of the puncture adapter
Remove the puncture adapter using the reverse procedure to that described in section 4-2 "Attaching of the
puncture adapter".
(1) Turn the knob counter clockwise until stopping, and remove the puncture adapter from the probe.
(2) Turn the knob counter clockwise.
Then remove the knob from the needle guide by pulling out the knob tip from the needle guide hole.
(3) Immediately wash and sterilize the puncture adapter (the needle guide and the knob), the needle stopper
and the depth gauge after it is removed from the probe.
4-4. Leaving the needle
The probe and the puncture adapter can be removed from the puncture needle during puncturing.
(1) Turn the knob counter clockwise until stopping, and remove the puncture adapter from the probe.
(2) Take the probe off the needle.
②
①
②
①
②
①