
-29-
MN1-1014 Rev. 14
Lock lever
Movable
range
○ mark
RELEASE
LOCK
RELEASE
LOCK
4-7. Removing from the ultrasound diagnostic instrument
The lock lever of the connector moves over the range shown in the figure at
right.
Align the
○
mark with the LOCK or RELEASE position and lock or release
the probe connector.
Use the procedure below to remove the probe from the probe connector.
• Removal procedure
The probe is removed when in one of the following states.
• The power switch is set to OFF.
• The image displayed on the ultrasound diagnostic instrument is
frozen
.
1. Turn the connector lock lever to align the ○ mark on the lever with the
RELEASE position.
2. Firmly grasp the connector unit and pull it out from the probe connector.
This completes the removal of the probe.
After use, perform washing, disinfection and sterilization of the probe
by following the procedure in section 5 “Washing, Disinfection and
Sterilization".
If the probe will not be used for an extended period of time, store it by
following the instructions in section 6 “Storage".
Содержание UST-9113P-3.5
Страница 2: ...MN1 1014 Rev 14 ii ...
Страница 16: ... 10 MN1 1014 Rev 14 2 Storage case Label A Label B ...
Страница 18: ... 12 MN1 1014 Rev 14 ...
Страница 24: ... 18 MN1 1014 Rev 14 ...
Страница 46: ... 40 MN1 1014 Rev 14 ...
Страница 48: ... 42 MN1 1014 Rev 14 ...
Страница 52: ... 46 MN1 1014 Rev 14 ...
Страница 54: ... 48 MN1 1014 Rev 14 ...