3-3-2. Checking the needle echo
(1) Dip the probe tip into the warm water so that the needle echo is displayed.
(2) Check the following points.
• The needle echo matches with the puncture guide line.
• The echo of the entire needle is displayed fully and clearly.
3-4. Performing washing and sterilization
(1) Before use, wash and sterilize the puncture adapter, the needle stopper and the depth gauge.
See section 5 "Washing, Disinfection and Sterilization"
(2) Wash and disinfect or sterilize the probe to be used in accordance with its usage purpose.
Warning
The equipment must be washed and sterilized before use.
Be sure to always properly wash and sterilize after use.
Failure to do so could result in an infection. Note that the equipment is not sterilized when shipped
from the factory. Before using the equipment for the first time, be sure to wash and sterilize it as
required.
Image on the monitor screen
Needle echo
Puncture guide line