
-23-
MN1-1014 Rev. 14
4. Usage
4-1. Operation
Bring the ultrasonic irradiation area of the probe to contact with the skin surface. An image of the region of
interest is displayed on the monitor of the ultrasound diagnostic instrument. For details on displaying and
adjusting the screens, see the documentation supplied with the ultrasound diagnostic instrument.
Caution
Do not move the probe with excessive force.
Pressing with more force than necessary can cause injury to the patient.
Scan for the minimum length of time necessary for the diagnosis and at the lowest suitable output.
There is the possibility that the patient’s internal tissues could be affected.
For details about the acoustic output, please refer to the documentation supplied with the ultrasound
diagnostic instrument.
Do not touch the connector terminal pin of the probe.
The probe may deteriorate or be damaged due to electrostatic discharge.
Do not touch the electronic probe connecting socket of the diagnostic instrument and the patient at
the same time.
It can cause electric shock to the patient.
Содержание UST-9113P-3.5
Страница 2: ...MN1 1014 Rev 14 ii ...
Страница 16: ... 10 MN1 1014 Rev 14 2 Storage case Label A Label B ...
Страница 18: ... 12 MN1 1014 Rev 14 ...
Страница 24: ... 18 MN1 1014 Rev 14 ...
Страница 46: ... 40 MN1 1014 Rev 14 ...
Страница 48: ... 42 MN1 1014 Rev 14 ...
Страница 52: ... 46 MN1 1014 Rev 14 ...
Страница 54: ... 48 MN1 1014 Rev 14 ...