The contact liquid is only renewed when the water level indicator lights up or there is visual contamination. If the AirOne Flex is not used for more than a week, please
remove the contact fluid and only refill the AirOne Flex when it is used again.
6.3 Further care instructions
Caution!
Do not carry out any descaling activities. These permanently damage the surface at the bottom of the nebulising chamber (ultrasonic oscillator / transducer)
and it will need to be replaced. Never touch or clean the oscillator with a hard object. Do not use acidic cleaning agents.
For optimal care of the nebulising cham-
ber, take a soft brush, pour water without additives into the nebulising chamber and use the brush to swirl up any possible residue and then rinse it out
.
You can wipe the housing with a damp cloth and then rub it dry
.
Any questions?
We’ll be happy to advise you - simply contact us on:
T: +49 (0)8086 933 327
T: +49 (0)5204 92970
For chargeable or warranty repairs, please download the Hippomed repair application and fill it out in full:
https://www.kerbl.com/de/service/downloads
7. Troubleshooting
Error
Remedy
Temperature LED lights up
Chapter 5.2.2.1
Water level LED lights up
Chapter 5.2.2.2
Horse is not inhaling properly
Check the inhalation valves for contamination
Check the tube connection between the unit and mask
Check the air supply setting on the mask
No inhalation mist visible
Check the content of the nebuliser cup
Unit not switched on (switch LED must light up green)
Oscillator contaminated - see Chapter 6.3
Oscillator faulty - contact the service centre
Unit does not switch on
No battery charge - charge the battery
Switch faulty - contact the service centre
Battery not charging/no LED display on the charger
Check the power supply to the charger
Battery not charging/LED display on the charger
See chapter 5.3 or the charging socket is oxidised: wet the plug with contact spray,
plug it in and out several times and turn it in the socket at the same time.
8. Warranty conditions
Albert Kerbl GmbH grants you a 2-year warranty on the ultrasound base unit in accordance with our general warranty conditions. These general warranty conditions can be
found at www.kerbl.com/guarantee. We will also be happy to send you the warranty conditions by post. Your statutory warranty rights remain unaffected by our guarantee.
This warranty begins on the date of purchase. The warranty does not cover:
1. Damage caused by natural wear and tear (wear parts such as the oscillator / transducer, battery), excessive strain and incorrect use (non-compliance with operating instruc-
tions).
2. Damage caused by force majeure or sustained during transport.
3. Damage to rubber parts, e.g. on access lines.
4. Damage to the mask including its attachments (valves) and the tube.
Warranty repairs are only carried out by the service centre for DE/AT or an authorised contracted workshop! The manufacturer reserves the right to reject any
warranty claims in the event of non-compliance.
The unit must for this purpose be delivered to the specified service address, at the risk of the unit's owner, complete, packaged and with sufficient postage, including the ori-
ginal invoice.
Please provide as much information as possible on the nature of the problem.