1.4 Measurement Workflow
22
1
Install the instrument, connect wires and cords, and
turn on the instrument.
1.4
Measurement Workflow
Installing the Instrument (p.5)
CHECKS
• Are the wires from the measurement targets shut off?
• Is the instrument turned off, and has the power cord been disconnected?
• Using D/A output (p.88)
• Using synchronized control to conduct measurements with 2
instruments simultaneously (p.81)
• Using external control to control integration (p.62)
• Sending and receiving data with the RS-232C, LAN, and GP-IB
interfaces (p.117)
Connecting Wires and Cords
Connect the measurement lines to the instrument and then connect the power cord to the instrument.
CHECKS
• Is the instrument connected on the secondary side of the circuit
breaker?
• Does the circuit being used exceed 1,000 V?
• Does the voltage or current being measured exceed 1,000 V or 70
A, respectively?
If so, use VT and CT.
• Are appropriate types of wire being used to connect to the voltage
and current input terminals?
Use solderless terminals that cover wiring with insulation. When
using solderless terminals with exposed metallic parts, use wire
with adequate dielectric strength and current capacity.
• Has the wiring been shorted?
• Are the input terminals loose?
• Have wires been connected properly?
Connect wires. (p.28)
e.g.)
Connect the power cord. (p.34)
When using one or more current sensors, see "3.9 Using a Current Sensor" (p.102).
Turning on the instrument (p.35)
Before turning on the instrument, verify that the wires have been connected properly one more time.
After displaying the initial screen, the instrument will display input values under the current settings.
Allow the instrument to warm up for at least 30 minutes.
Perform zero-adjustment.
To fulfill the instrument’s accuracy specifications, be sure to perform zero-adjustment for the voltage
and current measured values.
Содержание PW3336-01
Страница 2: ......
Страница 6: ...Contents iv...
Страница 16: ...Operating Precautions 10...
Страница 32: ...1 4 Measurement Workflow 26...
Страница 44: ...2 7 Turning Off the Instrument 38...
Страница 180: ...5 6 Wiring Specifications 174 3 3 phase 3 wire 3P3W 3P3W2M 4 3 phase 3 wire 3V3A 3P3W3M Current sensor...
Страница 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Страница 182: ...5 6 Wiring Specifications 176...
Страница 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Страница 200: ...Appendix 4 Dimensional Diagram A14...
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......