6.1 Troubleshooting
178
Before having the instrument repaired
When no apparent cause can be established
Perform a system reset.
This will return all settings to their factory defaults.
See:
3.10.4 Initializing the Instrument (System Reset) (p.112)
Symptom
Check Item, or Cause
Remedy and Reference
The display does not ap-
pear when you turn the
power on.
Has the power cord been disconnected
from the instrument? Is it connected
properly?
Verify that the power cord is connected properly.
See:
2.3 Connecting the Power Cord (p.34)
Keys do not work.
Has the key lock been activated?
Cancel the key lock state.
See:
3.10.3 Disabling Control Keys (Key Lock)
(p.111)
Is the REMOTE lamp lit up?
The instrument is being controlled by the control-
ler. End control and cancel the remote state.
See:
4.5 Canceling the Remote State (Activat-
ing the Local State) (p.134)
Содержание PW3336-01
Страница 2: ......
Страница 6: ...Contents iv...
Страница 16: ...Operating Precautions 10...
Страница 32: ...1 4 Measurement Workflow 26...
Страница 44: ...2 7 Turning Off the Instrument 38...
Страница 180: ...5 6 Wiring Specifications 174 3 3 phase 3 wire 3P3W 3P3W2M 4 3 phase 3 wire 3V3A 3P3W3M Current sensor...
Страница 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Страница 182: ...5 6 Wiring Specifications 176...
Страница 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Страница 200: ...Appendix 4 Dimensional Diagram A14...
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......