5.3 Measurement specifications
148
Voltage Ripple Rate Measurement Specifications
(Urf: Displayed as RF V % on panel display)
Current Ripple Rate Measurement Specifications
(Irf: Displayed as RF A % on panel display)
Harmonic Measurement Specifications
Measurement method
Calculates the AC component (peak to peak [peak width]) as a proportion of the voltage
DC component.
Effective measuring range
As per voltage and voltage waveform peak value effective measurement ranges.
Display range
0.00[%] to 500.00[%]
Polarity
None
Over-range display
Displays [o.r] when displaying [o.r] for the voltage waveform peak value and when the
voltage is [o.r] or 0.
Invalid data display
Urfsum is displayed as [- - - - -].
Measurement method
Calculates the AC component (peak to peak [peak width]) as a proportion of the current
DC component.
Effective measuring range
As per current and current waveform peak value effective measurement ranges.
Display range
0.00[%] to 500.00[%]
Polarity
None
Over-range display
Displays [o.r] when displaying [o.r] for the current waveform peak value and when the
current is [o.r] or 0.
Invalid data display
Irfsum is displayed as [- - - - -].
Measurement method
Zero-cross simultaneous calculation method (separate windows by channel according
to the wiring mode)
Uniform thinning between zero-cross events after processing with a digital antialiasing
filter
Interpolation calculations (Lagrange interpolation)
When the synchronization frequency falls within the 45 Hz to 66 Hz range
IEC 61000-4-7:2002 compliant
Gaps and overlaps may occur if the measurement frequency is not 50 Hz or 60 Hz.
When the synchronization frequency falls outside the 45 Hz to 66 Hz range
No gaps or overlap will occur.
Synchronization source
Conforms to synchronization source (SYNC) for the basic measurement specifications.
Measurement channels
3
Measurement items
Harmonic voltage RMS value, harmonic voltage content percentage, harmonic voltage
phase angle, harmonic current RMS value, harmonic current content percentage, har-
monic current phase angle, harmonic active power, harmonic active power content per-
centage, harmonic voltage/current phase difference, total harmonic voltage distortion,
total harmonic current distortion, fundamental wave voltage, fundamental wave current,
fundamental wave active power, fundamental wave apparent power, fundamental wave
reactive power, fundamental wave power factor, fundamental wave voltage current
phase difference, inter-channel voltage fundamental wave phase difference, inter-chan-
nel current fundamental wave phase difference
The following parameters can be downloaded as data with communications but not dis-
played:
Harmonic voltage phase angle, harmonic current phase angle, harmonic voltage cur-
rent phase difference
FFT processing word length
Number of FFT points
32 bits
4,096 points
Содержание PW3336-01
Страница 2: ......
Страница 6: ...Contents iv...
Страница 16: ...Operating Precautions 10...
Страница 32: ...1 4 Measurement Workflow 26...
Страница 44: ...2 7 Turning Off the Instrument 38...
Страница 180: ...5 6 Wiring Specifications 174 3 3 phase 3 wire 3P3W 3P3W2M 4 3 phase 3 wire 3V3A 3P3W3M Current sensor...
Страница 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Страница 182: ...5 6 Wiring Specifications 176...
Страница 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Страница 200: ...Appendix 4 Dimensional Diagram A14...
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......