![Hioki PW3336-01 Скачать руководство пользователя страница 152](http://html1.mh-extra.com/html/hioki/pw3336-01/pw3336-01_instruction-manual_2137351152.webp)
5.3 Measurement specifications
146
Current Waveform Peak Value Measurement Specifications
(Ipk: Displayed as A pk on panel display)
Efficiency Measurement Specifications (
η
: Displayed as
η
1 % or
η
2 % on display panel)
Measurement method
Measures the current waveform’s peak value (for both positive and negative polarity)
based on sampled instantaneous current values.
Range configuration
Measurement accuracy
±2.0% f.s. at DC and when 10 Hz
≤
f
≤
1 kHz (f.s.: current peak range). Provided as ref-
erence value when 0.1 Hz
≤
f < 10 Hz and when in excess of 1 kHz.
Effective measuring range
±5% to ±100% of current peak range (up to ±100 A)
Display range
±0.3% to ±102% of current peak range (values less than ±0.3% are subject to zero-sup-
pression)
Over-range display
Displays [o.r] when input exceeds ±102% of the current peak range.
Invalid data display
Ipksum is displayed as [- - - - -].
Current range
200 mA
500 mA
1 A
2 A
5 A
10 A
20 A
50 A
Current peak range
1.2000 A
3.0000 A
6.0000 A
12.000 A
30.000 A
60.000 A
120.00 A
300.00 A
Measurement method
Calculates the efficiency
η
[%] from the ratio of active power values for channels and wires.
Wiring modes and calculation
equations
Calculated based on the AC+DC rectifier active power.
Effective measuring range
As per the active power effective measurement range.
Display range
0.00[%] to 200.00[%]
Over-range display
Displays [o.r] when displaying [o.r] for the active power value being used as either the
numerator or denominator in the equation.
Displays [o.r] when the active power value being used as the denominator in the equation is 0.
Invalid data display
Displays [- - - - -] for wiring modes for which efficiency is not measured.
PW3336, PW3336-01, PW3336-02, PW3336-03
Wiring (WIRING)
CH1
CH2
Calculation formulas
1P2W
×
2
1P2W
1P2W
η
1=100
×
|P2|/|P1|
η
2=100
×
|P1|/|P2|
1P3W
1P3W
-
3P3W
3P3W
-
3P3W2M
3P3W2M
-
PW3337, PW3337-01, PW3337-02, PW3337-03
Wiring (WIRING)
CH1
CH2
CH3
Calculation formulas
1P2W
×
3
1P2W
1P2W
1P2W
η
1=100
×
|P3|/|P1|
η
2=100
×
|P1|/|P3|
1P3W&1P2W
1P3W
1P2W
η
1=100
×
|P3|/|Psum|
η
2=100
×
|Psum|/|P3|
3P3W&1P2W
3P3W
1P2W
3P3W2M
3P3W2M
-
3V3A
3V3A
-
3P3W3M
3P3W3M
-
3P4W
3P4W
-
Содержание PW3336-01
Страница 2: ......
Страница 6: ...Contents iv...
Страница 16: ...Operating Precautions 10...
Страница 32: ...1 4 Measurement Workflow 26...
Страница 44: ...2 7 Turning Off the Instrument 38...
Страница 180: ...5 6 Wiring Specifications 174 3 3 phase 3 wire 3P3W 3P3W2M 4 3 phase 3 wire 3V3A 3P3W3M Current sensor...
Страница 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Страница 182: ...5 6 Wiring Specifications 176...
Страница 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Страница 200: ...Appendix 4 Dimensional Diagram A14...
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......