4.2 Displaying PI and DAR
1
2
3
4
5
6
7
8
9
10
11
付録
索引
89
• The instrument automatically calculates and displays PI
(polarization index) and DAR (dielectric absorption ratio),
which are used as the criteria to determine the quality of
insulation.
Both measurements show a degree of change in insula-
tion resistance with time after a test voltage is applied.
Appendix 3 "Example of PI Criteria (Polarization Index)" (page 166)
• PI and DAR are calculated using the formulae below from
resistance values measured twice after a voltage is
applied. For PI, the measurement interval may be user-
set.
See 6.1 "Changing and Checking Interval Setting for PI Calculation"
4.2 Displaying PI and DAR
What is it used for?
Used to check whether insulation re-
sistance increases with time after a
voltage is applied.
(When the PI value or the DAR val-
ue is close to 1, the instrument de-
termines that the insulation of the
object under test is deteriorated.)
PI 10/1min
=
DAR 1min/15s =
DAR 1min/30s =
To determine DAR, press the
key to
turn off [
AVE
] on the screen before starting
measurement.
Resistance 10 min. after voltage application
Resistance 1 min. after voltage application
Resistance 1 min. after voltage application
Resistance 15 sec. after voltage application
Resistance 1 min. after voltage application
Resistance 30 sec. after voltage application
Содержание IR3455
Страница 2: ......
Страница 24: ...Operating Precautions 18...
Страница 110: ...4 4 Step Voltage Test 104...
Страница 160: ...7 3 9750 01 02 03 11 12 13 and 9751 01 02 03 ALLI GATOR CLIPs Specifications 154...
Страница 180: ...Appendix 4 Temperature Compensation Table 174...
Страница 181: ...1 2 3 4 5 6 7 8 9 10 11...
Страница 182: ......
Страница 183: ......
Страница 184: ......