7.2 Measurement Specifications
1
2
3
4
5
6
7
8
9
10
11
付録
索引
149
Temperature and humidity range for guaranteed accuracy:
0°C to 28°C, less than 80% RH (no condensation)
Measuring range
Measurement
accuracy
Equal to or less than the resistance
obtained by dividing the test voltage
(set value) by 100 nA.
±5%rdg.±5dgt.
Greater than the resistance calculated
by dividing the test voltage (setting
value) by 100 nA, less than or equal to
the resistance calculated by dividing
the test voltage (setting value) by 1
nA, or less than or equal to 500 G
±20%rdg.±5dgt.
Greater than the resistance calculated
by dividing the test voltage (setting
value) by 1 nA, or within the range of
501 G
to 9.99 T
±30%rdg.±50dgt.
Measurement Accuracy
Test voltage (set value)
Insulat
ion resi
stance
Содержание IR3455
Страница 2: ......
Страница 24: ...Operating Precautions 18...
Страница 110: ...4 4 Step Voltage Test 104...
Страница 160: ...7 3 9750 01 02 03 11 12 13 and 9751 01 02 03 ALLI GATOR CLIPs Specifications 154...
Страница 180: ...Appendix 4 Temperature Compensation Table 174...
Страница 181: ...1 2 3 4 5 6 7 8 9 10 11...
Страница 182: ......
Страница 183: ......
Страница 184: ......