3.2 Measuring Insulation Resistance
1
2
3
4
5
6
7
8
9
10
11
付録
索引
75
Every time
key is pressed, the indi-
cation changes in the order: resistance
current
DAR 1 min/15s
DAR 1 min/30s
PI
resistance
current
etc.
PI/DAR
See 4.2 "Displaying PI and DAR"
3.2.6 Insulation Resistance Measurement
Basis
When a high DC voltage is applied to an object under test, a
leakage current flows.
The insulation resistance instrument measures the applied
voltage V and the combined leakage current I and then cal-
culates the insulation resistance R.
Calculation formula R = V/I
I
C
and I
A
gradually decrease after the volt-
age is applied.
Holding data after measurement
If the indication is unstable, press the
key. The
average of the measurements is shown.
[
< 1.00 nA
] means "below 1.00 nA."