4.3 Temperature Compensation (TC)
96
Press the keys shown in the table below to switch the indi-
cation.
4.3.2 Exiting Temperature Compensation
Mode
Procedure
Press the
key.
[
TC
] is turned off and the instrument
exits temperature compensation mode.
• Resistance measured by the step voltage
test cannot be converted using
temperature compensation.
• In temperature compensation mode,
leakage current may be displayed by
pressing the
key but it cannot be
corrected for.
Indications to be switched
Keys used
key
Temperature
Reference temperature
Elapsed time
key
key
Insulation resistance
(after compensation)
Leakage current
(no compensation)
Setup screen of actual mea-
surement temperature
Standby state