7
Measuring Method
Measuring Method
Measurement Procedure
You will need (besides the device and
accessories
):
•
Model IM7580 series Impedance Analyzer × 1
•
Model IM9200 Test Fixture Stand × 1
•
Model IM9906 Adapter × 1
• 3.5 mm Connector torque wrench × 1 (This is not provided with the device.)
• Tweezers × 1 (This is not provided with the device. Tweezers which are thinner than the DUT
size, clean, and not magnetized are recommended.)
1
Connect the test head to the
impedance analyzer.
Connect the cable of the test head to the impedance analyzer.
See the instruction manual
of the impedance analyzer.
Test head
2
Attach the test head and adapter
to the stand.
Attach the test head and model IM9906 Adapter to the model
IM9200 Test Fixture Stand.
Model IM9200
Model IM9906
3
Execute the calibration.
Execute the open calibration, short calibration, and load
calibration.
4
Attach the device.
Attach the device to the model IM9200 Test Fixture Stand.
Device (model IM9201)
5
Execute the correction.
Execute the electrical length correction, short correction, and
open correction.
6
Start the measurement.
Connect the DUT to the measurement location of the device
guide and start the measurement by the impedance analyzer.
9
8
7
6
5
4
3
2
1
Appx.