18

Measuring Method

Open Correction

Open correction is executed to reduce errors. The GND plate is used.

You will need:

 

Hex driver × 1

 

• Tweezers × 1 (This is not provided with the device. Tweezers which are thinner than the DUT 

size, clean, and not magnetized are recommended.)

 

Device guide and GND plate × 1 of each

Device guides and GND plates for each DUT size

DUT size 

JIS (EIA)

Accessories

0603 (0201)

1005 (0402)

1608 (0603)

2012 (0805)

3225 (1210)
3216 (1206)

Device guide

0603

3225
3216

1005

1608

2012

0603

3225

3216

1005

1608

2012

0603

3225

3216

1005

1608

2012

0603

3225
3216

1005

1608

2012

0603

3225
3216

1005

1608

2012

GND plate
(At the time of open 
correction and 
measurement)

0603

3225
3216

1005

1608

2012

0603

3225

3216

1005

1608

2012

0603

3225

3216

1005

1608

2012

0603

3225
3216

1005

1608

2012

0603

3225
3216

1005

1608

2012

1

2

3

4

5

GND plate

Remove four bolts by using the hex driver.

Remove the device guide and short plate.

The plate might get deformed if it is removed forcibly. Keep 
the plate leveled so that load is not applied over it.

Attach the GND plate to the electrode.

The frame of DUT size to be measured should fit the 

measurement position.

Attach the device guide.

The frame of DUT size to be measured should fit the 

measurement position.

Tighten four bolts by using the hex driver.

6

Keep the pusher in a raised state.

Содержание IM9201

Страница 1: ...EN Instruction Manual IM9201 SMD TEST FIXTURE July 2016 Edition 1 IM9201A961 00 16 07H...

Страница 2: ......

Страница 3: ...impedance analyzer carefully Contents Verifying Package Contents p 2 Safety Information p 3 Operating Precautions p 4 Product Overview p 5 Names and Functions of Parts p 6 Measuring Method p 7 Measur...

Страница 4: ...4 Knob for fixing tip pin 1 Tip pin 2 3 Brush for cleaning 1 4 Hex driver 1 5 Carrying case 1 6 Instruction Manual 1 GND plate and device guide for the sizes 1005 0402 1608 0603 2012 0805 3216 1206 32...

Страница 5: ...WARNING Indicates a potentially hazardous situation that may result in death or serious injury to the operator CAUTION Indicates a potentially hazardous situation that may result in minor or moderate...

Страница 6: ...not press the tip pin in the electrode when the DUT is not attached While pressing the tip pin into the DUT the DUT may pop out To avoid injuries lower the pusher slowly and press the tip pin into th...

Страница 7: ...5 2 5 0 15 H 0 4 L W H Tip pins device guides and plates for each DUT size DUT size JIS EIA Accessories 0603 0201 1005 0402 1608 0603 2012 0805 3216 1206 3225 1210 Tip pin 3225 3216 1005 1608 2012 322...

Страница 8: ...lacing the GND plate and DUT on top of the electrode 7 Model IM9906 Adapter 3 5 mm 7 mm The exclusive adapter for connecting the device to the test head of the impedance analyzer side 8 Test head The...

Страница 9: ...er See the instruction manual of the impedance analyzer Test head 2 Attach the test head and adapter to the stand Attach the test head and model IM9906 Adapter to the model IM9200 Test Fixture Stand M...

Страница 10: ...lution Is the short plate worn Compare the resistance value with the new or a non defective short plate Replace with new short plate if the resistance value is high Is the tip pin deformed If the tip...

Страница 11: ...d Model IM9200 1 2 Metal fittings 1 Pull the metal fittings to the model IM9200 Test Fixture Stand side 2 Place the test head on the stand and tighten the knob Model IM9906 Notch Measuring terminal 3...

Страница 12: ...ay cause damage to the electrode When the calibration cannot be executed Confirm that the calibration kit and model IM9906 Adapter are connected properly Check that the DC measurement limit value of t...

Страница 13: ...ration by using the impedance analyzer 2 1 0 0 3 4 Nut 7 Short calibration kit 9 7 Remove the open calibration kit by turning the nut of the open calibration kit 8 Place the short calibration kit on t...

Страница 14: ...only the nut of load calibration kit while pressing the top part of the load calibration kit If the top part of the load calibration kit is rotated the internal conductors of the calibration kit and a...

Страница 15: ...ower the bolt thread by turning the nut of model IM9906 Adapter 3 2 2 2 Loosen the stand fixing knob 3 Place the device on the model IM9200 Test Fixture Stand As shown in figure slide the notch part 4...

Страница 16: ...Correction The error caused by the phase shift occurring between the calibration reference surface and the measurement DUT s connection surface is compensated Set the electrical length correction to...

Страница 17: ...ility You will need Hex driver 1 Tweezers 1 This is not provided with the device Tweezers which are thinner than the DUT size clean and not magnetized are recommended Device guide and short plate 1 ea...

Страница 18: ...lts However measurement error may increase due to variability in the short plate s inductance 2 3 4 5 6 Short plate 2 Remove four bolts by using the hex driver 3 Remove the device guide and GND plate...

Страница 19: ...tip pin is finalized fix it by turning the knob and lower the pusher 12 Execute short correction by using the impedance analyzer When the correction cannot be executed Confirm that the short calibrati...

Страница 20: ...608 2012 0603 3225 3216 1005 1608 2012 GND plate At the time of open correction and measurement 0603 3225 3216 1005 1608 2012 0603 3225 3216 1005 1608 2012 0603 3225 3216 1005 1608 2012 0603 3225 3216...

Страница 21: ...hod 7 Start the open correction in the impedance analyzer 1 2 3 4 5 6 When correction cannot be executed check that the DC measurement limit value of the impedance analyzer is appropriate 9 8 7 6 5 4...

Страница 22: ...t location remove it 2 2 Pick up the DUT with tweezers and set it to the measurement location of device guide Keep the surface with electrode facing down Set the DUT perpendicular to the frame of the...

Страница 23: ...mm 1608 0603 1 6 0 15 0 8 0 1 mm H 0 4 mm 2012 0805 2 0 0 15 1 25 0 15 mm H 0 4 mm 3216 1206 3 2 0 15 1 6 0 15 mm H 0 4 mm 3225 1210 3 2 0 15 2 5 0 15 mm H 0 4 mm The measurement unit of size 3225 121...

Страница 24: ...parts used in the product may deteriorate with extended use To ensure the product can be used over the long term it is recommended to replace these parts on a periodic basis When replacing parts plea...

Страница 25: ...16 01 EN...

Страница 26: ......

Страница 27: ......

Страница 28: ......

Отзывы: