Measuring Method
Short Correction
Short correction is executed to reduce errors. The short plate is used.
CAUTION
• To avoid damage to the device and DUT, while pressing the tip pin into the DUT, move
the tip pin slowly so that shock is not applied.
• To avoid damage to the bolt fixing the plates, insert the bolt vertically and tighten it
with appropriate force.
• To avoid injury, make sure that your hand is not in-between the pusher and tip pin.
IMPORTANT
Do not scratch the electrode, GND plate, and short plate of the device or touch them directly with
your hand. It may affect the measurement accuracy and reproducibility.
You will need:
•
Hex driver × 1
• Tweezers × 1 (This is not provided with the device. Tweezers which are thinner than the DUT
size, clean, and not magnetized are recommended.)
•
Device guide and short plate × 1 each
Device guides and short plates for each DUT size
The set direction as viewed from the front (position from where the model name IM9201 can be read).
DUT size
JIS (EIA)
Accessories
0603 (0201)
1005 (0402)
1608 (0603)
2012 (0805)
3225 (1210)
3216 (1206)
Device guide
0603
3225
3216
1005
1608
2012
0603
3225
3216
1005
1608
2012
0603
3225
3216
1005
1608
2012
0603
3225
3216
1005
1608
2012
0603
3225
3216
1005
1608
2012
Short plate
(At the time of
short correction)
0603
3225
3216
1005
1608
2012
0603
3225
3216
1005
1608
2012
0603
3225
3216
1005
1608
2012
0603
3225
3216
1005
1608
2012
0603
3225
3216
1005
1608
2012
9
8
7
6
5
4
3
2
1
Appx.
Index
Содержание IM9201
Страница 1: ...EN Instruction Manual IM9201 SMD TEST FIXTURE July 2016 Edition 1 IM9201A961 00 16 07H...
Страница 2: ......
Страница 25: ...16 01 EN...
Страница 26: ......
Страница 27: ......
Страница 28: ......