9

Measuring Method

Attaching the Test Head and the Adapter

Attach the test head and model IM9906 Adapter to the model IM9200 Test Fixture Stand.

IMPORTANT

The impedance analyzer, test head, and measurement cable are adjusted together at the time 

of shipment. Use the test head and measurement cable included along with the impedance 

analyzer. Confirm that the serial no. of the impedance analyzer and test head is same.

You will need:

 

Test head × 1

 

Model IM9200 Test Fixture Stand × 1

 

Model IM9906 Adapter × 1

 

• 3.5 mm Connector torque wrench × 1 (This is not provided with the device.)

Test head

Model IM9200

1

2

Metal fittings

Pull the metal fittings to the model IM9200 Test 

Fixture Stand side.

Place the test head on the stand and tighten the 
knob.

Model IM9906

Notch

Measuring 
terminal

3

4

Place the notch of the model IM9906 Adapter 

parallel to the metal fittings, then install onto the 

measurement terminal of the test head.

At this time it is stopped temporarily. Position the notch of the 

model IM9906 Adapter where the metal fittings can slide.

Fix the notch of model IM9906 adapter by sliding 

the metal fittings.

3.5 mm connector 
nut

5

Tighten the nut of the 3.5 mm connector by using a 
torque wrench.

Recommended torque: 0.9 N

·

m

Tightening the nut until the handle of the wrench bends 

slightly is sufficient. Do not over tighten.

9

8

7

6

5

4

3

2

1

Appx.

Index

Содержание IM9201

Страница 1: ...EN Instruction Manual IM9201 SMD TEST FIXTURE July 2016 Edition 1 IM9201A961 00 16 07H...

Страница 2: ......

Страница 3: ...impedance analyzer carefully Contents Verifying Package Contents p 2 Safety Information p 3 Operating Precautions p 4 Product Overview p 5 Names and Functions of Parts p 6 Measuring Method p 7 Measur...

Страница 4: ...4 Knob for fixing tip pin 1 Tip pin 2 3 Brush for cleaning 1 4 Hex driver 1 5 Carrying case 1 6 Instruction Manual 1 GND plate and device guide for the sizes 1005 0402 1608 0603 2012 0805 3216 1206 32...

Страница 5: ...WARNING Indicates a potentially hazardous situation that may result in death or serious injury to the operator CAUTION Indicates a potentially hazardous situation that may result in minor or moderate...

Страница 6: ...not press the tip pin in the electrode when the DUT is not attached While pressing the tip pin into the DUT the DUT may pop out To avoid injuries lower the pusher slowly and press the tip pin into th...

Страница 7: ...5 2 5 0 15 H 0 4 L W H Tip pins device guides and plates for each DUT size DUT size JIS EIA Accessories 0603 0201 1005 0402 1608 0603 2012 0805 3216 1206 3225 1210 Tip pin 3225 3216 1005 1608 2012 322...

Страница 8: ...lacing the GND plate and DUT on top of the electrode 7 Model IM9906 Adapter 3 5 mm 7 mm The exclusive adapter for connecting the device to the test head of the impedance analyzer side 8 Test head The...

Страница 9: ...er See the instruction manual of the impedance analyzer Test head 2 Attach the test head and adapter to the stand Attach the test head and model IM9906 Adapter to the model IM9200 Test Fixture Stand M...

Страница 10: ...lution Is the short plate worn Compare the resistance value with the new or a non defective short plate Replace with new short plate if the resistance value is high Is the tip pin deformed If the tip...

Страница 11: ...d Model IM9200 1 2 Metal fittings 1 Pull the metal fittings to the model IM9200 Test Fixture Stand side 2 Place the test head on the stand and tighten the knob Model IM9906 Notch Measuring terminal 3...

Страница 12: ...ay cause damage to the electrode When the calibration cannot be executed Confirm that the calibration kit and model IM9906 Adapter are connected properly Check that the DC measurement limit value of t...

Страница 13: ...ration by using the impedance analyzer 2 1 0 0 3 4 Nut 7 Short calibration kit 9 7 Remove the open calibration kit by turning the nut of the open calibration kit 8 Place the short calibration kit on t...

Страница 14: ...only the nut of load calibration kit while pressing the top part of the load calibration kit If the top part of the load calibration kit is rotated the internal conductors of the calibration kit and a...

Страница 15: ...ower the bolt thread by turning the nut of model IM9906 Adapter 3 2 2 2 Loosen the stand fixing knob 3 Place the device on the model IM9200 Test Fixture Stand As shown in figure slide the notch part 4...

Страница 16: ...Correction The error caused by the phase shift occurring between the calibration reference surface and the measurement DUT s connection surface is compensated Set the electrical length correction to...

Страница 17: ...ility You will need Hex driver 1 Tweezers 1 This is not provided with the device Tweezers which are thinner than the DUT size clean and not magnetized are recommended Device guide and short plate 1 ea...

Страница 18: ...lts However measurement error may increase due to variability in the short plate s inductance 2 3 4 5 6 Short plate 2 Remove four bolts by using the hex driver 3 Remove the device guide and GND plate...

Страница 19: ...tip pin is finalized fix it by turning the knob and lower the pusher 12 Execute short correction by using the impedance analyzer When the correction cannot be executed Confirm that the short calibrati...

Страница 20: ...608 2012 0603 3225 3216 1005 1608 2012 GND plate At the time of open correction and measurement 0603 3225 3216 1005 1608 2012 0603 3225 3216 1005 1608 2012 0603 3225 3216 1005 1608 2012 0603 3225 3216...

Страница 21: ...hod 7 Start the open correction in the impedance analyzer 1 2 3 4 5 6 When correction cannot be executed check that the DC measurement limit value of the impedance analyzer is appropriate 9 8 7 6 5 4...

Страница 22: ...t location remove it 2 2 Pick up the DUT with tweezers and set it to the measurement location of device guide Keep the surface with electrode facing down Set the DUT perpendicular to the frame of the...

Страница 23: ...mm 1608 0603 1 6 0 15 0 8 0 1 mm H 0 4 mm 2012 0805 2 0 0 15 1 25 0 15 mm H 0 4 mm 3216 1206 3 2 0 15 1 6 0 15 mm H 0 4 mm 3225 1210 3 2 0 15 2 5 0 15 mm H 0 4 mm The measurement unit of size 3225 121...

Страница 24: ...parts used in the product may deteriorate with extended use To ensure the product can be used over the long term it is recommended to replace these parts on a periodic basis When replacing parts plea...

Страница 25: ...16 01 EN...

Страница 26: ......

Страница 27: ......

Страница 28: ......

Отзывы: