
11
Measuring Method
Nut
5
Open
calibration kit
4
Place the open calibration kit on the reference
surface.
5
As shown in figure, connect by turning only the nut
of open calibration kit while pressing the convex
portion on the top part of the open calibration kit.
The convex part of the open calibration kit is connected with
the internal conductor. If the calibration kit is rotated without
pressing the convex part, the internal conductors of calibration
kit and adapter may graze each other, and the calibration kit
and adapter may get damaged.
6
Execute open calibration by using the impedance
analyzer.
2
1
3
4
Nut
7
Short
calibration kit
9
7
Remove the open calibration kit by turning the nut of
the open calibration kit.
8
Place the short calibration kit on the reference
surface.
9
As shown in figure, connect by turning only the nut
of short calibration kit while pressing the convex
portion on the top part of the short calibration kit.
The convex part of the short calibration kit is connected with
the internal conductor. If the calibration kit is rotated without
pressing the convex part, the internal conductors of calibration
kit and adapter may graze each other, and the calibration kit
and adapter may get damaged.
10
Execute short calibration by using the impedance
analyzer.
2
1
3
4
9
8
7
6
5
4
3
2
1
Appx.
Index
Содержание IM9201
Страница 1: ...EN Instruction Manual IM9201 SMD TEST FIXTURE July 2016 Edition 1 IM9201A961 00 16 07H...
Страница 2: ......
Страница 25: ...16 01 EN...
Страница 26: ......
Страница 27: ......
Страница 28: ......