12
Measuring Method
11
Load
calibration kit
13
Nut
11
Remove the short calibration kit by turning the nut of
the open calibration kit.
12
Place the load calibration kit on the reference
surface.
13
As shown in figure, connect by turning only the nut
of load calibration kit while pressing the top part of
the load calibration kit.
If the top part of the load calibration kit is rotated, the internal
conductors of the calibration kit and adapter may graze each
other, and the calibration kit and adapter may get damaged.
14
Execute load calibration by using the impedance
analyzer.
2
1
3
4
15
Remove the load calibration kit by turning the nut of
the load calibration kit.