20
Measuring Method
Starting the Measurement
Place the DUT in the measurement location of the device guide, lower the pusher and start the
measurement.
You will need:
• Tweezers × 1 (This is not provided with the device. Tweezers which are thinner than the DUT
size, clean and not magnetized are recommended.)
•
DUT × 1
Measurement location of device guide
0603
3225
3216
1005
1608
2012
0603 (0201)
0603
3225
3216
1005
1608
2012
3216 (1206)
3225 (1210)
2012
(0805)
0603
3225
3216
1005
1608
2012
1005
(0402)
1608 (0603)
JIS (EIA)
1
If there is a DUT in the measurement location,
remove it.
2
2
Pick up the DUT with tweezers and set it to the
measurement location of device guide.
Keep the surface with electrode facing down. Set the DUT
perpendicular to the frame of the device guide such that
the DUT and electrode come in contact with each other.
Stopper
Pusher
3
4
3
Hold the pusher of the stopper side tightly.
4
Pull the stopper slowly.
5
5
Lower the pusher slowly and press the DUT with
the tip pin.
If the DUT pops out
•
If there is a deviation in the setting of the DUT or
position of the tip pin, reset the DUT and adjust the
position of the tip pin.
• If the pusher suddenly drops down, then lower the
pusher slowly next time.
6
Start the measurement using the impedance
analyzer.