80
Table 7.3 Setting Control Program Messages 6/10
Mnemonic Description
Format
V C M
Selection of auto voltage check execute mode
d1 (Select: 0 ~ 1)
0: OFF
1: ON
[Format]
VCM d1
d1: NR1 format
V C M ?
Query response of auto voltage monitoring
execute mode is same as setting.
[Format]
VCM?
[Response]
d1
C C M
Selection of auto contact check execute mode
d1 (Select: 0 ~ 1)
0: OFF
1: ON
[Format]
CCM d1
d1: NR1 format
C C M ?
Query response of auto contact check execute
mode is same as setting.
[Format]
CCM?
[Response]
d1
W C P
Work capacity setting
Capacity of work for open correction value
calculation
d1 (0.5 ~ 99.9) pF
[Format]
WCP d1
d1: NR2 format
W C P ?
Query response of work capacity is same as
setting.
[Format]
WCP?
[Response]
d1
O S T ?
Executes open correction once and capacity
value measured during open correction is
returned as a response.
In case of error, “999.9” is returned.
d1 (Jig capacity: 0 ~ 99.9)
In error: 999.9
[Note] Be sure to carry out open correction once
before performing a contact check.
[Format]
OST?
[Response]
d1: NR2 format
V C K ?
Executes voltage check once and returns result
of it as a response.
d1 (0 ~ 1)
0: NO 1: GO
[Format]
VCK?
[Response]
d1
C C K ?
Executes contact check once and returns result
of it as a response.
d1 (0 ~ 1)
0: NO 1: GO
[Format]
CCK?
[Response]
d1
Содержание DSM-8104
Страница 2: ......
Страница 20: ...10 3 2 Block Diagram...
Страница 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...
Страница 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...
Страница 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...
Страница 130: ...120 11 Product Full View...
Страница 131: ......
Страница 132: ......