37
5.1.11 Setting operating environment
Press the [SETUP] (SET UP) key on the Measurement screen.
The screen changes to the Operating Environment Setting screen. (See Fig. 5.8)
9.9999E+99Ω
M E S . V : 0 . 1 V A V E : O F F
S A M P L : 1 0 0 m S S E Q : O F F : 0
D E L A Y : 0 m S R A N G E : A U T O 1
C O N F S E L F C A L P O W R
Fig. 5.8 Operating Environment Setting Screen
(1) Self-calibration setting
Press the [F3] CAL (CALIBRATION) key on the Operating Environment Setting
screen.
The screen changes to the Self-calibration Setting screen. (See Fig. 5.9)
C A L I B R A T I O N
A U T O M O D E : O F F / O N
I N T E R V A L : [ 6 0 ] S e c
D O W N U P E X E C
Fig. 5.9 Self-calibration Setting Screen
The function keys will change to [F2] DOWN, [F3] UP and [F5] EXEC.
Select an item to be set by operating the [
▲
] and [
▼
] keys.
After finishing settings, press the [ENTER] key to return to the Measurement
screen.
(2) “AUTO MODE” setting
This function executes self-calibration (self-calibration check of the A/D
converter and range) automatically at a preset time interval.
OFF: Auto self-calibration is not executed.
ON: Auto self-calibration is executed.
(3) “INTERVAL” setting
Set a time interval for auto self-calibration by inputting a time interval operating
the numeric input keys and pressing the [ENTER] key.
Or, a time interval can be selected from among designated values operating the
[F2] DOWN and [F3] UP keys because specified values are already set in the
meter.
Содержание DSM-8104
Страница 2: ......
Страница 20: ...10 3 2 Block Diagram...
Страница 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...
Страница 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...
Страница 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...
Страница 130: ...120 11 Product Full View...
Страница 131: ......
Страница 132: ......