6
2.2.8 Storage and display functions of measured data
(1) Measured data buffer
Measured data up to 1000 data groups can be sequentially stored.
Up to 1000 most recent data groups are stored and displayed on the
measurement screen.
Note: Stored data is cleared when the power is shut down.
(2) Histogram counter
Measured values are classified into ten types and numbers of data groups are
stored by digitally counting them by a counter. Thresholds are set for the
individual measuring modes and are indicated on the screen in a bar graph.
Thresholds are set within the measuring range.
Note: Stored data is cleared when the power is shut down.
2.2.9 Operability and display
(1) Keyboard
Rubber key switches
Key clicking tone on or off can be selected
Key lock (prevention of malfunction in the Remote mode) on or off can be
selected
(2) Displays
1) Liquid crystal display 240x64dot graphic LCD (30 columns, 8 rows)
Backlight (Yellowish green LEDs)
Character size
Measured result
4 x 2 size
Measuring conditions
Full size
Bar graph display
2) High-voltage warning display Red LED lit if higher than about 30V
2.2.10 Resume function
When the power supply goes off, the set parameters are automatically stored. When the
power is turned on again, the operation is resumed at the set parameters that are
automatically stored, except the voltage impression state.
2.2.11 Input and output functions (external control interfaces)
(1) GP-IB interface
All items that are operated on the panel can be GP-IB controlled.
Measured results comprising 5-digit exponential measured values, comparison
results and check of measured state are sent in one line of data strings.
Responses in 4-digit fixed decimal points or 5-digit integers are sent to query
commands.
(2) Specification of RS-232 interface function
All items that are operated on the panel can be RS-232 controlled.
Measured results comprising 5-digit exponential measured values, comparison
results and check of measured state are sent in one line of data strings.
Responses in 4-digit fixed decimal points or 5-digit integers are sent to query
commands.
Содержание DSM-8104
Страница 2: ......
Страница 20: ...10 3 2 Block Diagram...
Страница 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...
Страница 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...
Страница 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...
Страница 130: ...120 11 Product Full View...
Страница 131: ......
Страница 132: ......