7.13  Status Byte Register 
 
The Status Byte register is the core of the service request function and all status 
information and event information is concentrated in this register. 
The following messages relate to the Status Byte register. 
 

*CLS 

 

         

 

         

 

         

 

         

 

Clears the following registers: 

· Status Byte register 
· Standard event status register 
· Device event status register 
· Error register 

*SRE 

  Sets Service Request Enable register 

*SRE?

  Queries Service Request Enable register 

*STB?

  Queries Status Byte Register 

 
 

 
 
 

 

ERR : ERROR 

 

 

RQS : Request Service 

 

  MSS : Master Summary Status 

 

 

ESB : Event Summary Bit   

  MAV : Message Available 

 

 

DSE : Device Event Summary Bit   MEC : Measure Complete 

 

Fig. 7.5 Configuration of Status Byte Register 

 

RQS 

 

 

Service Rquest 

Generation 

ERR 

ESB

MAV

DSB

 

 

MEC

Status Byte 

Register

  

MSS 

 

 

 

 

 

 

 

Logical

 

OR 

 

 

ERR 

 

ESB

MAV

DSB

 

 

MEC

Service Request 

Enable Register

 

Содержание DSM-8104

Страница 1: ...DIGITAL SUPER INSTRUCTION MANUAL DSM 8104 April 2013 Revised edition 1 DS8104A981 01 13 04H MEGOHMMETER...

Страница 2: ......

Страница 3: ...eloped to measure insulating materials of a high resistance value capable of measuring resistance within the range of 1 x 103 to 3 x 1016 and a current within the range of 30fA to 10mA The meter is su...

Страница 4: ...described in this manual under the following headings Strictly follow these precautions and operational instructions Warning Indicates a potentially hazardous situation that could result in an electr...

Страница 5: ...nted on the front and rear panels 5 Preparations for Measurement This chapter describes the screen of the meter for measurement settings on the screen operating method methods for indicating measured...

Страница 6: ...bustible gas exists Otherwise an explosion or a fire may break out Be certain to ground the grounding pin of the power cable to prevent an accident If the grounding pin of the power cable cannot be gr...

Страница 7: ...self diagnosis function 5 2 2 5 Comparison measurement and deviation percent measurement functions 5 2 2 6 Measurement jig setting and data processing function 5 2 2 7 Measuring sequence program 5 2...

Страница 8: ...5 2 3 Deviation display 43 5 2 4 Creating histogram 44 6 Measurement 6 1 Functions of and Connecting Measuring Terminals 47 6 1 1 Functions of measuring terminals 47 6 1 2 Connecting measuring termina...

Страница 9: ...7 13 Status Byte Register 91 7 14 Using Program Message 93 7 15 Standard Event Status Register 96 7 16 Error Register 98 7 17 Device Event Status Register 100 7 18 Initialization Value by RST Message...

Страница 10: ...116 10 2 1 Equipment needed for calibration 116 10 2 2 Calibration of measuring voltage 116 10 2 3 Calibration of current measurement 116 10 3 Self calibration Function 117 10 4 Self diagnosis Functi...

Страница 11: ...nual 1 This manual If any damage or missing accessory is found please contact our distributor or sales office near your location 1 2 Operating Power Supply This meter is powered by the following power...

Страница 12: ...length of measuring cable Depending on the length of the measuring cable adjustment of the meter is necessary to correct open values in the contact check function The meter was adjusted for use of the...

Страница 13: ...300ms 2 Temperature range 23 5 C humidity 85 RH or less 3 Mode Self calibration automatically executed at 1 minute interval 4 Averaging process On Remarks The resistance measurement accuracy greatly...

Страница 14: ...OFF by a setting 2 Current setting errors are 10 2 2 Functional Specification 2 2 1 Measuring time Item Setting Range Delay Time 0 9999 ms Averaging Cycles Auto setting Sampling time setting power pe...

Страница 15: ...limit comparison HI GO Measured value upper limit value Intermediate comparison IN GO Upper limit value measured value lower limit value Lower limit comparison LO GO Upper limit value measured value C...

Страница 16: ...cter size Measured result 4 x 2 size Measuring conditions Full size Bar graph display 2 High voltage warning display Red LED lit if higher than about 30V 2 2 10 Resume function When the power supply g...

Страница 17: ...ement end measurement computing end troubles that have occurred Electrical specification Contact input and output by a photocoupler Connector product No is 57RE 40360 730B DDK 2 3 General Specificatio...

Страница 18: ...er SME 8310 With changeover switch for surface and volume measurement and interlock Electrode for planar sample 19 6 diameter SME 8311 Same as above Weight electrode SME 8320 For surface and volume re...

Страница 19: ...capable of outputting 1000V 10mA and 250V 50mA maximum supplying stable measuring voltage The measuring voltage source of this meter can set a large maximum output current so that measuring time can...

Страница 20: ...10 3 2 Block Diagram...

Страница 21: ...nd rear panels of this meter Warning symbol This symbol is marked where the operational manual needs to be referred to Electrical This symbol indicates high voltage hazard and is indicated on terminal...

Страница 22: ...ng of trigger input 4 START lamp Pressing the START key lights up this lamp going off when the STOP key is pressed The set voltage is output and trigger input can be received while the START lamp is l...

Страница 23: ...on to display measured results measuring conditions and various setting screens 12 INPUT connector The connector for measurement input The connector of a double structure comprising a center conductor...

Страница 24: ...on 21 HANDLER connector The connector for handler connection 22 INTERLOCK connector The input connector for connection of signals from a fixture when an interlock function is used 23 EXT TRIGGER conne...

Страница 25: ...setting the measurement screen and the setting screen by pressing this key LOCK 0 Key lock Press this key to prohibit key input Pressing this key prohibits operations of the keys other than the STOP a...

Страница 26: ...check The key to execute once a contact check When in the Numeric Input mode the key will become a BS back space key to delete characters that are input MOD 4 Deviation Measuring mode The key to move...

Страница 27: ...ential Measurement monitoring screen When in the Numeric Input mode the key will become an input key 7 COMP 8 Compare The key to move to the Comparison Measurement Setting screen When in the Numeric I...

Страница 28: ...ide mark will be lit when the number of data groups saved by the buffering function for 1000 data groups exceeds 1000 Measured data after this mark is lit will be discarded without being stored in the...

Страница 29: ...unction is not on when the side mark is not displayed The side mark flashes when a decision result is NG V CHK ON A side mark on the screen indicates the ON OFF status of the voltage check function Th...

Страница 30: ...nt thanks to the resume function Warning Before turning the power on check that the AC power supply voltage to be used matches the power supply voltage specified on the rear panel If they do not match...

Страница 31: ...In addition to a Self diagnosis function this meter has a Self calibration function which allows users to execute self calibration checks at preset intervals The factory default settings are Self cali...

Страница 32: ...ng for key input Waiting for key input display Waiting for key input Waiting for key input Waiting for key input Waiting for key input Waiting for key input Waiting for key input Waiting for key input...

Страница 33: ...OCK INTERLOCK BUF FULL CHARGE ON OPEN SET MEAS COND SHIFT INT TRIG MAN EXT COMP ON V CHK ON C CHK ON Regular Measurement screen Function key display Press the MONI key for changeover Sequential Measur...

Страница 34: ...y during Auto Range will mean that the value is beyond the measurable range Table 5 3 Meanings of Side Marks Panel Character Side Mark Status REMOTE In external remote mode KEY LOCK In key lock mode I...

Страница 35: ...tting erase a numeric value by pressing the BS key and press the ENTER key to return to data before the setting The functions keys corresponding to each setting item will be the scroll keys F2 F3 for...

Страница 36: ...100 nF 4 10 nF 5 1 nF 6 100 pF 7 10 pF 8 10 pF 2 RANGE Current range F2 DOWN F3 UP F4 AUTO F5 HOLD SEQ Sequential measurement Program No 0 9 ON OFF F2 DOWN F3 UP F4 ON F5 OFF DCHG1 Discharging time Be...

Страница 37: ...ressing the ENTER key Move to the setting screen operating setting and function keys Cannot measure on this screen Fig 5 4 Key Operation and Screen Change Table 5 5 Setting keys and function keys for...

Страница 38: ...rrent measurement A Surface resistivity measurement Rs Volume resistivity measurement Rv Notice The actual unit for surface resistivity measurement is The actual unit for volume resistivity measuremen...

Страница 39: ...e contact for protection of the relay contact This procedure is also repeated when changing the CHARGE terminal for precharging Protective resistance value measuring voltage maximum permissible curren...

Страница 40: ...ing on the filter response time Select the high speed changeover mode with the filter turned off in measurement that requires fast changeover of measuring voltage The low noise mode with the filter tu...

Страница 41: ...ling time Stable high sensitivity current measurement is sometimes disabled due to impacts of noise components contained in the input current This meter performs A D conversion after integrating the i...

Страница 42: ...the F5 PSET key to set the Measuring Condition Setting screen On the Measuring Condition Setting screen the F4 and F5 keys change to keys that select ON and OFF respectively b Move to AVE operating t...

Страница 43: ...under the same conditions 1 Discharge 1 Set a discharging time before impressing voltage 2 Charge Set a time for impressing voltage before starting measurement 3 Measurement Set a time for measuremen...

Страница 44: ...program to be created Ten types of sequence programs 0 to 9 can be created 3 Charging time measuring time and discharging time setting In sequence program measurement the operational sequence of DISC...

Страница 45: ...P and F5 END In the absence of data in the data buffer none will be displayed The content in the data buffer will be 0 0000E 00 in the resistance measuring mode includes volume and surface resistiviti...

Страница 46: ...is useful such as when analyzing data after taking measurement Measurement can be performed relatively quickly if data is fetched collectively after finishing all measurements since fetching of data f...

Страница 47: ...EXEC Select an item to be set by operating the and keys After finishing settings press the ENTER key to return to the Measurement screen 2 AUTO MODE setting This function executes self calibration se...

Страница 48: ...t 2s when auto execute is on 2 Auto self calibration cannot be performed even when auto execute is set to ON if sequential measurement is ON 5 1 12 Setting other items The following items are set Inte...

Страница 49: ...f open circuited high level input When disabled the START key will be disabled and a side mark will light up in the INTERLOCK display on the left of the LCD screen CUTOFF Disables the interlock functi...

Страница 50: ...en to display measured results and measuring conditions As shown in Fig 5 3 the Measured Value Display screen is split into the Regular Measurement screen and Sequential Measurement Monitoring screen...

Страница 51: ...s The following menu is displayed for the function keys Functions UNIT EXP FIG Measured value is displayed in unit F1 UNIT UNIT Measured value is displayed in EXPONEN F2 EXP EXPONENT Number of display...

Страница 52: ...mode by keying the and keys Select HI IN or LO Setting of UPPER or LOWER sets the input status of the numeric keys Input a numeric value to be input Refer to 1 and 2 in the following Press the ENTER k...

Страница 53: ...n flashes if the decision result of a comparison measurement is NG If beep tone output is set a beep tone will be emitted as a decision result and the decision result will be output to the connector o...

Страница 54: ...external trigger and sequential measurement This function cannot be used in the Internal Trigger mode 1 Press the DATA MEASURED RESULT DATA key Pressing the key changes the screen to the screen to sel...

Страница 55: ...he position desired to change and input a new value The setting in that position will be changed Press the ENTER key to set the input threshold When setting is finished press the ENTER key to return t...

Страница 56: ...ntity larger than Threshold 1 the number shown in 2 represents a data quantity between Thresholds 2 and 1 and similarly the numbers shown in represents data group quantities smaller than Threshold 9 T...

Страница 57: ...grounding terminal for reducing impacts of noise and hazard of an electrical shock Normally connected to the OUTPUT terminal or GUARD terminal by a short bar OUTPUT A measuring voltage output termina...

Страница 58: ...inal voltage must be impressed to the terminal on the grounding side when insulation resistance insulation resistance to the earth of a sample a part or circuit whose one wire is grounded such as a li...

Страница 59: ...part or material and by providing a guard the same potential is obtained between the guard circuit and ground if the GUARD terminal is connected to the GROUND terminal Warning Connecting the GUARD ter...

Страница 60: ...insulation resistance is measured by measuring a current that flows through a sample and flows from one terminal to a terminal on the other side and by converting the current into a resistance value...

Страница 61: ...ed by the equivalent circuit shown in Fig 6 4 a without a guard The equivalent circuit shows that Current i1 flowing through the sample and current i2 that flows through Insulating Materials a and b b...

Страница 62: ...ults can be set b Output a trigger pulse on timing to change a sample and input the pulse to the EXT TRIGGER connector or the HANDLER connector on the rear panel of the meter In this event the meter s...

Страница 63: ...an be shortened and measurement accuracy can be enhanced by taking measurement after charging a sample in advance when automatically measuring a sample of a large electrostatic capacity such as a capa...

Страница 64: ...ly a dielectric absorption current is substantially large compared with a leakage current and time constant of it is very large depending on the type of insulating material so that a long time is need...

Страница 65: ...evices of a structure that become terminals only by contacting a sample are used more Terminals for measurement must be free of malcontact and must contact the sample tightly However the shape and str...

Страница 66: ...ances of insulating materials are mutually compared insulation resistances unique to insulating materials that are not affected by terminal shape or other factors are needed These are volume resistivi...

Страница 67: ...g mode The meter is capable of automatically performing calculations of volume resistivity and surface resistivity mentioned in 1 For this purpose set in advance constants of electrodes to be used in...

Страница 68: ...ode diameter t Specified value of sample thickness If D1 26 D2 38 If D1 50 D2 70 If D1 70 D2 90 If D1 76 D2 88 Press the ENTER key after finishing setting The screen will return to the measurement scr...

Страница 69: ...in 6 2 Measuring Parts and Circuits and 6 3 Measuring Planar Sample when a shielding box is used 2 Interlock function setting Output of measuring voltage is sometimes not safe such as when the lid of...

Страница 70: ...and a side mark will light up by INTERLOCK indication on the left of the LCD screen CUTOFF The interlock function will be disabled Measuring voltage can always be output irrespective of the INTERLOCK...

Страница 71: ...sistivity Set an electrode constant as follows On the screen illustrated in Fig 6 6 press the F5 ACTL key The screen will change to the Electrode Constant Setting screen illustrated in Fig 6 10 Input...

Страница 72: ...Be sue to use a shielded conductor cable as the measuring lead connected to the INPUT terminal to prevent induced troubles such as noise and connect the shielded conductor to the GUARD terminal Conne...

Страница 73: ...ecessary by operating the panel keys or by issuing commands through interfaces Notice Measurement and data output will be made as in normal operation even if a decision of voltage check is NG However...

Страница 74: ...reen referring to 4 5 Display Screen and 5 1 1 Screen types Select on the screen whether to execute not execute contact check during measurement Selection made for contact checks will be indicated by...

Страница 75: ...capacity value of the measuring system 3 Setting decision capacity reference value A minimum capacity value of a sample is set operating the numeric keys 0 5 to 99 9 The minimum capacity value is 1 10...

Страница 76: ...ses 0 to 30 can be set Remote status cancel Cancelled by pressing LOCAL key on the panel Table 7 1 Interface Functions Function Description SH1 Source handshake All functions AH1 Acceptor handshake Al...

Страница 77: ...output bit0 Voltage check NG Result of auto execution bit1 Contact check NG Result of auto execution bit2 Overrange If status is 0 results are normal in all events Comparison result 0 HI Measured val...

Страница 78: ...the following formats NR1 NR2 NR3 any ASCII character string format and binary format with a specified length See Table 7 3 List of Program Messages when deciding which format will be used NR1 format...

Страница 79: ...wing three delimiters by the DLM command EOI Default EOI EOI Note A default will result when the power is turned on 7 5 Listener Function The DSM 8104 accepts program messages in ASCII character strin...

Страница 80: ...ers CR LF EOI simultaneously with LF LF EOI simultaneously with LF CR EOI simultaneously with CR EOI simultaneously with last data byte Format of command program message Commands with plural program d...

Страница 81: ...ys are disabled except the LOCAL STOP and LCDOF keys Press the LOCAL key to return to the Local mode The LOCAL key is also disabled when the Local Lockout LLO command is executed by an interface The S...

Страница 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...

Страница 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...

Страница 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...

Страница 85: ...data format designation d1 Format designation 0 3 0 Basic format Default 1 Measured values only output 2 Comparison results only output 3 No data output Returns to default when power is turned on Form...

Страница 86: ...0 1 0 OFF Lamp extinguished 1 ON Lamp lit Corresponds to LCDOF key Format DCD d1 d1 NR1 format Query response in LCD display mode is same as setting Format LCD Response d1 Setting number of effective...

Страница 87: ...it set values d1 Unit 0 1 0 PLC 1 ms d2 Set value PLC 1 15 ms 2 300 Format SPL d1 d2 d1 NR1 format d2 NR1 format Query response on setting number of integral time unit set value is same as setting For...

Страница 88: ...Format Measuring voltage setting d1 Voltage 0 1 1000 0 V Format IVS d1 d1 NR2 format Query response of measuring voltage is same as setting 11 Format IVS Response d1 Measuring power source setting d1...

Страница 89: ...at DEV Response d1 d2 Electrode data setting d1 SIZE ACTL selection 0 ACTUAL 1 SIZE d2 Electrode inside diameter SIZE 0 0 999 9mm d3 Electrode outside diameter SIZE 0 1 1199 9mm d4 Sample thickness SI...

Страница 90: ...ng Capacity of work for open correction value calculation d1 0 5 99 9 pF Format WCP d1 d1 NR2 format Query response of work capacity is same as setting Format WCP Response d1 Executes open correction...

Страница 91: ...rnal Trigger mode Needs not be used in manual trigger external trigger and sequential measurement Format RDT d1 d1 NR1 format Response See 7 3 Comparison Measuring mode setting d1 Comparison execute 0...

Страница 92: ...NR2 format d4 NR2 format d5 NR2 format d6 NR2 format Same as set items of query SEQ command message in sequential mode Returns current set data d1 Mode 0 1 d2 Program No 0 9 d3 Discharging time before...

Страница 93: ...2d d3 d4 d5 d6 d7 d8 d9 Histogram counter readout Returns count values of bar graphs divided into ten and displayed on the screen as responses d1 to d2 are output in order of large threshold Format RH...

Страница 94: ...on current range calibration setting d1 Auto self calibration done and not done 0 1 0 OFF 1 ON d2 Auto self calibration interval time 10 9999s Format ACL d1 d2 d1 NR1 format d2 NR1 format Query respon...

Страница 95: ...he device event status enable register d1 0 255 See Fig 7 8 and Table 7 8 Note Data will be cleared simultaneously with response output Format DSR Response d1 NR1 format Environmental data save d1 Env...

Страница 96: ...Status register clear See Fig 7 5 Format CLS Sets the service request enable register d1 0 255 See Fig 7 5 and Table 7 5 Format SRE d1 d1 NR1 format Query of data in the service request enable registe...

Страница 97: ...pment inspection at the factory Shuts down if in start state Format RST Sets OPC bit of the standard event status register after all operations being executed are finished This command is used for det...

Страница 98: ...message execute The following command messages can be executed only in a start state If executed in a stop state no processing will be performed and a result of the previous execution will return VCK...

Страница 99: ...the enable register masks enable or disable for each event or status before ORing them As the MSS Master Summary Status bit the bit in the DIO7 position becomes logical OR of the seven other bits Thes...

Страница 100: ...URQ 6 CME 5 EXE 4 DDE 3 QYE 2 RQC 1 OPC 0 7 6 BOV 5 BFL 4 STP 3 ITL 2 LM2 1 LM1 0 Device Event Status Register Logical OR 7 6 BOV 5 BFL 4 STP 3 ITL 2 LM2 1 LM1 0 Device Event Status Enable Register Q...

Страница 101: ...ter Device event status register Error register Sets Service Request Enable register Queries Service Request Enable register Queries Status Byte Register ERR ERROR RQS Request Service MSS Master Summa...

Страница 102: ...are set to enable has been set RQS is cleared by SR function status byte read during serial polling but MSS is not cleared Standard event has occurred One of the bits of the standard event status regi...

Страница 103: ...valid unless this condition is met 5 THL command Histogram counter threshold setting Do not omit program data by this command Other commands that have more than one program data groups can omit unnece...

Страница 104: ...ommands and trigger commands have response data and detection of end of processing of them can most suitably be accomplished by monitoring the MAV bit of the Status Byte register The MAV bit remains s...

Страница 105: ...n continually executing commands When executing commands that have response data send the commands in one line by separating them by as shown earlier 3 Necessity for detection of end of command proces...

Страница 106: ...er Sets the Standard Event Status Enable register Queries the Standard Event Status Enable register Queries and clears the Standard Event Status register Standard Event Standard Register Standard Even...

Страница 107: ...will be ignored in execution Execution error A received message cannot be executed at present or a set parameter value is outside the range These are errors grouped in DRE and CNE bits of the Error r...

Страница 108: ...ing registers Status Byte register Standard Event Status register Device Event Status register Error register Queries and clears the Error register MLE Message Length Error HDE Header Error DFE Data F...

Страница 109: ...of parameters has exceeded a specified number There are parameters that cannot be recognized Reset by reading this register Data range error Set when a parameter value is outside specified range Rese...

Страница 110: ...ent Status register Device Event Status register Error register Sets Device Event Status Enable register Queries Device Event Status Enable register Queries Device Event Status register Device Event S...

Страница 111: ...ter Measured data buffer full Set when the measured data buffer becomes full Reset when the buffer becomes empty Measurement stop event Set by one of the following factors The STOP key is pressed The...

Страница 112: ...RST Measuring mode MOD Resistance measurement Resistance measurement Trigger mode TGM Internal Internal Measuring voltage IVS 0 1 V 0 1 V Integral time SPL 300 ms 300 ms Averaging AVE ON ON Measuring...

Страница 113: ...ER t THICKNESS K ELC 50 0 mm 70 0 mm 0 1 mm 0 01 Interfaces INTERLOCK BEEP ON OFF BEEP NO GO CLICK GP IB ADRS 232 BAND 232 DATA 232 PARI 232 STOP CNF CUTOFF ON NO ON 1 4800 7BIT NON 1BIT Auto calibrat...

Страница 114: ...Nos and Signal Names Pin No Signal Name Function Remarks Reception carrier detection Not used with DSM 8104 Reception data Transmission data Data terminal ready Signal ground Data setting ready Not u...

Страница 115: ...he XON XOFF code Notice Data is sometimes lost if flow control by a control signal is not performed Wait a suitable time when processing a command message without a response A suitable time must be co...

Страница 116: ...state from a remote state Table 8 3 Invalid Messages Mnemonic Description OPC bit setting when processing ends 1 is returned when operation completes Sending a command message requesting a response in...

Страница 117: ...GGER Input V CHECK Input C CHECK Input OPEN Input START Input STOP Input INTERLOCK Input KEYLOCK Input V CHECK GO Output C CHECK GO Output OPEN GO Output COMP GO Output EOM Output INDEX Output SRT E O...

Страница 118: ...ce when sequential measurement is ON STOP Stop input and has the same function as that of the STOP key on the panel Turns off voltage output and stops measurement Stop input can always be input Not di...

Страница 119: ...ut for acceptable non acceptable decision in comparison measurement Output becomes active when decision is acceptable as measurement result EOM Output signal to indicate that measured data can be outp...

Страница 120: ...ommon terminal of the output signals is DC isolated from the internal circuit An equivalent circuit of the output circuit is illustrated in Fig 9 1 Output Signal Characteristics Output Voltage LOW HIG...

Страница 121: ...lent circuit of the input circuit is illustrated in Fig 9 2 Input Characteristics Maximum Input Voltage LOW HIGH Maximum Input Current 0 5 V 5 V 1 2 mA Notice Output the output of the handler side by...

Страница 122: ...nnector outer conductor is DC isolated from the internal circuit The common side is connected to the input signal common terminal Pins 2 and 20 of the handler interface An equivalent circuit of the in...

Страница 123: ...connected to the 5V power source inside Output the control side output by an open collector or TTL output The common side connector outer conductor of the interlock signal is DC isolated from the int...

Страница 124: ...ular measurement Internal trigger mode INDEX EOM V CHECK GO C CHECK GO COMP GO 3 Regular measurement Sequence off External Trigger mode TRIGGER INDEX EOM V CHECK GO C CHECK GO COMP GO 4 Open correctio...

Страница 125: ...check only V CHECK INDEX EOM V CHECK GO 6 Execution of contact check only C CHECK INDEX EOM C CHECK GO 7 Sequential measurement START VON INDEX EOM PROGRAM V CHECK GO C CHECK GO COMP GO Discharge Cha...

Страница 126: ...esistance value of a sample based on measuring voltage impressed to the sample and on measured current value The resistance measurement accuracy of this meter is calibrated by calibrating the output v...

Страница 127: ...asuring voltage of 1V The relationship between the reference resistance and current measuring range is shown below Measuring Voltage Reference Resistance Current Range HOLD1 HOLD2 HOLD3 HOLD4 HOLD5 HO...

Страница 128: ...y a function that monitors internal operations the meter stops its operation and displays the following error message screen In this case a component of the meter is faulty Please contact the HIOKI sa...

Страница 129: ...n 85 RH covering it with sheeting to prevent dust and dirt 2 Moving and transporting When moving or transporting the meter please exercise caution to avoid shocks and vibration to the meter When trans...

Страница 130: ...120 11 Product Full View...

Страница 131: ......

Страница 132: ......

Отзывы: