98
7.16 Error
Register
The Error register controls error information and is an 8-bit register.
The data of this register is grouped in the CME, EXE, DDE and QYE bits of the Standard
Event Status register. (Mask processing is not performed)
The following messages relate to the Error register.
*CLS
Clears the following registers:
· Status Byte register
· Standard Event Status register
· Device Event Status register
· Error register
ERR?
Queries and clears the Error register
MLE : Message Length Error
HDE : Header Error
DFE : Data Format Error
DRE : Data Range Error
CNE : Can Not Execute
REQ : Read Empty Queue
QOF : Queue Overflow
ISE : Internal communication Error
Fig. 7.7 Configuration of Error Register
7
MLE
6
HDE
5
DFE
4
DRE
3
CNE
2
ISE
1
BDE
0
Error Register
PON
7
URQ
6
CME
5
EXE
4
DDE
3
QYE
2
RQC
1
OPC
0
Standard Event
Status Register