109
(2) Output signals
OUTCOM:
The common terminal for output signals.
Isolated from the common terminal for input signals.
/V.CHECK GO: Output for “acceptable/non-acceptable” decision in voltage check.
Output becomes active when decision is “acceptable.”
/C.CHECK GO: Output for “acceptable/non-acceptable” decision in contact check.
Output becomes active when decision is “acceptable.”
/OPEN GO:
Output for “acceptable/non-acceptable” decision in open
correction.
Output becomes active when decision is “acceptable.”
/COMP GO:
Output for “acceptable/non-acceptable” decision in comparison
measurement.
Output becomes active when decision is “acceptable” as
measurement result.
/EOM:
Output signal to indicate that measured data can be output.
Fetch output data on the timing this signal becomes LOW level.
/INDEX:
Output signal to indicate that a sample can be changed.
Change the sample on the timing this signal becomes LOW level.
/SRT.E:
Signal indicating that the meter can be started (voltage can be
output).
Becomes HIGH level in a start disable state by the interlock
function, becoming LOW level when starting is ready.
/VON:
Signal to indicate measuring voltage is being output.
/ERROR:
Output signal indicating a trouble has occurred with the meter.
Содержание DSM-8104
Страница 2: ......
Страница 20: ...10 3 2 Block Diagram...
Страница 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...
Страница 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...
Страница 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...
Страница 130: ...120 11 Product Full View...
Страница 131: ......
Страница 132: ......