73
────────────────────────────────────────────────────
────────────────────────────────────────────────────
Application
Fitted in a 3194 product (factory-fitted option)
Measurement lines
Single-phase,two-wire (1P2W)/ Single-phase,three-wire (1P3W)/ Three-
phase,three-wire (3V3A, 3P3W)/ Three-phase,four-wire (3P4W)
Number of channels
Maximum of 3 channels selectable from channels 1 to 6, depending on 3194
product wiring mode
1ch
2ch
3ch
4ch
5ch
6ch
Using channels
①
1P2W
1P2W
1P2W
1P2W
1P2W
1P2W 1+2+3, 2+3+4, 3+4+5, 4+5+6
②
1P3W/3P3W
1P2W
1P2W
1P2W
1P2W 12+3, 3+4+5, 4+5+6
③
1P3W/3P3W
1P3W/3P3W
1P2W
1P2W 12+3, 34+5, 4+5+6
④
1P3W/3P3W
1P3W/3P3W
1P3W/3P3W
12+3, 34+5, 56
⑤
3V3A/3P4W
1P2W
1P2W
1P2W 123, 4+5+6
⑥
3V3A/3P4W
1P3W/3P3W
1P2W 123, 45+6
⑦
3V3A/3P4W
3V3A/3P4W
123, 456
Measurement range
Basic frequency 10 Hz to 4.5 kHz
Measurement system
PLL synchronization/fixed clock
Analysis method
FFT
Type of window
Rectangular (gaps in window)
Display update rate
Every 1 window (excluding when FD/printer output and communication)
A/D
12 bits
Computational accuracy
32 bits (floating-point calculations)
PLL source
Selectable from the following sources:
U of the selected combination of channels for measurement
I of the selected combination of channels for measurement
Derived from an external sync signal
External Sync Signal
Possible to synchronize with external synchronization signals input from the
3194 main product's analog connector for U/I/P analysis.
Input level (for sine wave, 1 to 10 Vrms, 100 kHz or less)
Frequency splitting function included (1/1 to 1/255)
Crest factor
2.5 max. (current, voltage)
Output function
FD, printer, GP-IB, RS-232C
Chapter 9
Specifications
(1) General Specifications
Содержание 9605-01
Страница 1: ...HARMONIC INSTRUCTION MANUAL 9605 01 MEASUREMENTS UNIT...
Страница 2: ......
Страница 10: ...4 2 2 Screen Configuration...
Страница 42: ...36...
Страница 48: ...42 6 2 Timer Control of Output...
Страница 78: ...72 8 7 The Output Item Bits...