3.6 Zero Adjustment Operation
────────────────────────────────────────────────────
NOTE
3.6 Zero Adjustment Operation
Zero adjustment must be carried out after warming up.
1. Press the
STATUS
key, then use the
PAGE
keys
to move the cursor to "
HARM
," to display the
harmonic measurement setting screen.
2. Move the cursor to "
WIRING.
"
3. Press the
F5
"
ZERO ADj
" key to start zero
adjustment.
・
The operation is carried out for 3 channels simultaneously.
・
The operation takes about 3 seconds.
・
Carry out the operation with no inputs on either voltage or current.
・
Carry out the operation after degaussing (DMAG) when using the 9600 or
9602.