8.3 Command Reference
────────────────────────────────────────────────────
Syntax
:DATAout:ITEM:HARMonic:ALLClear
Example
":DATAout:ITEM:HARMonic:ALLClear"
Function
Clears all output items set by the
":DATAout:ITEM:HARMonic"
command.
Syntax
:DATAout:ITEM:HARMonic:ORDer
<NR1>,<NR1>,<ODD/EVEN/ALL>
first <NR1> = 0 to 50 (lower limit
order)
second <NR1> = 0 to 50 (upper limit
order)
Example
":DATAout:ITEM:HARMonic:ORDer
1,15,ODD"
Sets the output harmonic order to 1st
to 15th odd-order.
Function
Sets the output item for the harmonic
order (level, percentage, and phase
angle) to FDD or printer.
Syntax
:DATAout:ITEM:HARMonic:ORDer?
Response
syntax
":DATAOUT:ITEM:HARMONIC:ORDER
<0-50>,<0-50>,<ODD/EVEN/ALL>
"
Example
Transmission
Response
":DATAout:ITEM:HARMonic:ORDer?"
":DATAOUT:ITEM:HARMONIC:ORDER
1,15,ODD"
Function
Queries the output item for the
harmonic order (level, percentage, and
phase angle) to FDD or printer.
8
.3.2 Specific Commands for Harmonic Analysis Function
:DATAout:ITEM:HARMonic:ALLClear
Clears all output items for harmonic default.
:DATAout:ITEM:HARMonic:ORDer
Sets the output item for the harmonic order.
:DATAout:ITEM:HARMonic:ORDer?
Queries the output order of the harmonic data.