62
────────────────────────────────────────────────────
8.3 Command Reference
────────────────────────────────────────────────────
Syntax
:MEASure:ITEM:HARMonic:NORMal
<NR1>,..(up to 5 items)
<NR1> = 0 to 63
Example
":MEASure:ITEM:HARMonic:NORMal
9,1,9,9,0"
As the default output items for the
normal measurement, H
U1, HI1, HP1,
HTRU1, HTRI1, HTFU1, HTFI1
are
specified.
Function
Sets the default items (rms value, active
power, total harmonic distortion ratio) to
be transferred in the response message
to the ":MEASure:HARMonic?" query
in the default mode.
The item is set as shown below by
setting bits, to specify a single
numerical value.
Note
If the setting value is out of range, an
execution error occurs.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
RMS value data1
-
-
HI3
HI2
HI1
HU3
HU2
HU1
Power
data2
-
-
-
-
HPSUM
HP3
HP2
HP1
THD-R
data3
-
-
HTRI3 HTRI2 HTRI1 HTRU3 HTRU2 HTRU1
THD-F
data4
-
-
HTFI3 HTFI2 HTFI1 HTFU3 HTFU2 HTFU1
Frequency data5
-
-
-
-
-
-
-
HF
Syntax
MEASure:ITEM:HARMonic:NORMal?
Response
syntax
":MEASURE:ITEM:HARMONIC:NORMAL
<0-63>,..(up to 5 items)
"
Example
Transmission
Response
":MEASure:ITEM:HARMonic:NORMal?"
":MEASURE:ITEM:HARMONIC:NORMAL
9,1,9,9,0"
Function
Queries the setting items specified by
the ":MEASure:ITEM:HARMonic:
NORMal" command.
:MEASure:ITEM:HARMonic:NORMal
Sets the output item for the harmonic measurement value.
:MEASure:ITEM:HARMonic:NORMal?
Queries the output item for the harmonic measurement value.
Содержание 9605-01
Страница 1: ...HARMONIC INSTRUCTION MANUAL 9605 01 MEASUREMENTS UNIT...
Страница 2: ......
Страница 10: ...4 2 2 Screen Configuration...
Страница 42: ...36...
Страница 48: ...42 6 2 Timer Control of Output...
Страница 78: ...72 8 7 The Output Item Bits...