6
────────────────────────────────────────────────────
3.2 Analysis Channel Selection
────────────────────────────────────────────────────
NOTE
Wiring mode and selected channels
1ch
2ch
3ch
4ch
5ch
6ch
Using channels
①
1P2W
1P2W
1P2W
1P2W
1P2W
1P2W
1+2+3, 2+3+4, 3+4+5, 4+5+6
②
1P3W/3P3W
1P2W
1P2W
1P2W
1P2W
12+3, 3+4+5, 4+5+6
③
1P3W/3P3W
1P3W/3P3W
1P2W
1P2W
12+3, 34+5, 4+5+6
④
1P3W/3P3W
1P3W/3P3W
1P3W/3P3W
12+3, 34+5, 56
⑤
3V3A/3P4W
1P2W
1P2W
1P2W
123, 4+5+6
⑥
3V3A/3P4W
1P3W/3P3W
1P2W
123, 45+6
⑦
3V3A/3P4W
3V3A/3P4W
123, 456
3.2 Analysis Channel Selection
You can select three input unit channels for harmonic analysis from the
input units (maximum three channels) in the 3194. If the frequencies are the
same, analysis on three single-phase channels is possible.
1. Press the
STATUS
key, then use the
PAGE
keys
to move the cursor to "
HARM
," to display the
harmonic measurement setting screen. This
displays the selected wiring mode.
2. Use the
CURSOR
keys to select the "
WIRING
"
item.
3. Move the cursor to the channel to be analyzed
with the
CURSOR
keys, then press the
F1
"
SET
"
key to complete the setting.
・
The channel selection is always of adjacent input units, corresponding to the
wiring modes set on the 3194 product. If the 3194 product wiring mode is
set to 3P3W for each of channels 1 and 2 and channels 3 and 4, then the
selection becomes channels 1, 2, and 3. In this case, the third channel is
treated as single phase for analysis.
・
Up to three channels can be selected; a single-channel analysis specification
is not possible.