1.3 Name and Function of the Parts
20
POWER LED
Goes on or flashes when power is supplied
to the module.
Remains on, flashes, or changes to another
color according to the state of the module.
POWER LED indication
Lit in green
: Data being recorded.
Flashing in green : Standing by.
Lit in yellow
: Alarm output.
Flashing in yellow : It indicates one of the
following:
• The voltage is outside the effective mea-
surement range.
• The current is out of range.
• The active power is a negative value.
Lit in red
: Non-recoverable error
occurred.
*1
Flashing in red
: Recoverable error
occurred.
*2
Mark area
Use this area to make a note of the object
to measure or the module ID.
Use an ink pen, since pencil lead may rub
off.
CLAMP SENSOR
INPUT terminals
Connect the output of clamp sensors to
these terminals (for 3 channels).
U INPUT
terminals
Connect voltages to be measured to these
terminals.
ALARM terminals
Connect the alarm output cable to these
terminals.
These terminals are electrically insulated
from the input terminals.
U RANGE switch
Select a voltage range of 100 V (70 to 130
VAC) or 200 V (140 to 260 VAC).
Module ID setting
dial
Use the dial to set the module's identifica-
tion No.
*1: The module needs repair. Contact your vendor (agent) or
nearest Hioki office.
*2: The same module ID may be used by another module.
Содержание 2331-20
Страница 1: ...INSTRUCTION MANUAL POWER METER 2331 20 MODULE ...
Страница 2: ......
Страница 25: ...21 1 4 Dimension Diagrams 1 4 Dimension Diagrams 94 5 1 103 3 1 103 9 1 110 1 Unit mm 96 1 45 5 1 4 0 5 ...
Страница 26: ...1 4 Dimension Diagrams 22 ...
Страница 28: ...2 1 Setting the Module ID 24 ...
Страница 44: ...4 2 Insulation of Internal Circuit 40 ...
Страница 54: ...5 3 General Specifications 50 ...
Страница 56: ...6 2 Service 52 ...
Страница 57: ......
Страница 58: ......