Notes on Use
Follow these precautions to ensure safe operation
and to obtain the full benefits of the various func-
tions.
Notes on Use
• The maximum input voltage is as follows.
U INPUT: 300 Vrms, 424.3 V peak
ALARM : 30 VDC
CLAMP SENSOR INPUT: 1.5 Vrms, 2.2 V
peak
Attempting to measure voltage in excess
of the maximum input could destroy the
instrument and result in personal injury or
death.
• The maximum rated voltage between input
terminals and ground is 300 Vrms.
Attempting to measure voltages exceeding
300 Vrms with respect to ground could
damage the instrument and result in
personal injury.
Do not allow the instrument to get wet, and do
not take measurements with wet hands.
This may cause an electric shock.
Do not use the instrument where it may be
exposed to corrosive or combustible gases.
The instrument may be damaged or cause an
explosion.
the instrument
to get wet.
Do not allow
Corrosive or
combustible
gases