Chapter 4: Software Interface
52
Exicor 150AT User Manual
•
Birefringence:
the magnitude of linear retardation normalized to the beam path
length in the sample (sample thickness) – units are nanometers per centimeter
(nm/cm).
•
Stress
: the estimated magnitude of mechanical stress – units are pounds per
square inch (psi) or mega pascals MPa. This requires a stress optic coefficient.
•
DC
: the strength of the detector signal – units are volts (V). This is useful for
investigating any possible correlation between retardation magnitude and signal
strength. Measurements can sometimes be affected if the beam is clipping an
edge of the sample or the sample holder (this condition reduces the signal
strength at the detector).
The
Graph Type
control selects the method used to display the data. The options
are
•
Mag
: 2D color plot of one of the measurements above.
•
Angle
: 2D vector plot of only the fast axis angle.
•
Mag & Angle
: 2D color plot of the magnitude of one of the measurement
types above with the angle vector plot overlaid on top.
•
Mag 3D
: 3D surface representation of the magnitude of the one of the
measurement types above. The magnitude is represented both by color and
height of the surface.
•
Line Chart
: Line graph view of the data points in the order collected. All
measurement types are shown together.
Legend
The legend shows the color key for the currently selected plot. There is a drop-down
menu that allows the user to select the type of legend for the plot. The types of
legends are:
•
Custom
: Data points are shown as a specific color. The magnitude range is
divided into 10 sections, each having a unique color. The legend minimum,
maximum, and color for each section is customizable.
•
Gray Scale
: Data points are shown as shades of gray. The selected range
is shown on a continuous spectrum from black to white. Black represents low
numerical results while white represents high numeric results. The legend
minimum and maximum values are customizable.
•
Spectral
: Data points are shown as shades of the visible color spectrum.
The selected range is shown on a continuous spectrum from
violet
to
red
.
Violet
represents low numerical results while
red
represents high numeric
results. The legend minimum and maximum values are customizable.
Figure 4.15 below shows the 3 different types of legends shown together in one
figure.
Содержание EXICOR 150AT
Страница 1: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F ...
Страница 2: ......
Страница 3: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F Software Version 6 5 ...
Страница 7: ...Table Of Contents Exicor 150AT User Manual v ...
Страница 16: ...Operator s Safety Summary Exicor 150AT User Manual 6 Figure S 1 Laser Beam Path ...
Страница 17: ...Laser Safety Exicor 150AT User Manual 7 Figure S 2 Compliance Labeling ...
Страница 21: ...Exicor 150AT User Manual 11 ...
Страница 26: ...Chapter 1 Overview of the Exicor Birefringence Measurement System 16 Exicor 150AT User Manual ...
Страница 29: ...Loading a Sample Exicor 150AT User Manual 19 ...
Страница 49: ...Configuring the System Exicor 150AT User Manual 39 Figure 4 6 Auto Calibration Settings Dialog Box ...
Страница 74: ...Chapter 4 Operating the System 64 Exicor 150AT User Manual ...
Страница 81: ...Cleaning the Cabinet Exicor 150AT User Manual 71 ...
Страница 90: ...Appendix F Exicor 150AT Limited Warranty 80 Exicor 150AT User Manual ...