Making Birefringence Measurements
Exicor 150AT User Manual
49
SETTING/
BEHAVIOR
MODE
Regular Speed
High Speed
LIA
Sensitivity
As defined in
Configuration
System Parameters
HLIAs
LIA Time
Constant
As defined in
Configuration
System Parameters
Settings
“4 ms” Fixed
Number of
Averages
As defined in
Configuration
System Parameters
Settings
Fixed to 1
Measurement
Repeatability
Higher
- Depends on LIA
sensitivity and number of
averages
Lower
– LIA sensitivity and
number of averages is set to
speed-optimized settings
Measurement
Speed
Slower
– Measures as fast as it
can given the number of
averages and the need to update
the data plot
Faster
– Fixed at precisely 100
points/second for stationary
scans
Plot Update
Real Time
At End of Scan
Table 4.4 Overview of High Speed Mode vs. Regular Speed Mode
Defining a Zone Scan
The “zones” scan type divides the scan area into 4 areas so that the same scan can
be performed in each area and the data for each area saved to a separate file. This
feature is only compatible with a zones sample holder. A zones scan is defined by
the zone numbers to scan and the area to scan within each zone. Zone scans are
rectangular shapes only.
Follow these steps to define a zones scan:
1. Select
Zones
from the drop-down menu under
Scan Region
. The scan
control now looks like Figure 4.13 below.
Содержание EXICOR 150AT
Страница 1: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F ...
Страница 2: ......
Страница 3: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F Software Version 6 5 ...
Страница 7: ...Table Of Contents Exicor 150AT User Manual v ...
Страница 16: ...Operator s Safety Summary Exicor 150AT User Manual 6 Figure S 1 Laser Beam Path ...
Страница 17: ...Laser Safety Exicor 150AT User Manual 7 Figure S 2 Compliance Labeling ...
Страница 21: ...Exicor 150AT User Manual 11 ...
Страница 26: ...Chapter 1 Overview of the Exicor Birefringence Measurement System 16 Exicor 150AT User Manual ...
Страница 29: ...Loading a Sample Exicor 150AT User Manual 19 ...
Страница 49: ...Configuring the System Exicor 150AT User Manual 39 Figure 4 6 Auto Calibration Settings Dialog Box ...
Страница 74: ...Chapter 4 Operating the System 64 Exicor 150AT User Manual ...
Страница 81: ...Cleaning the Cabinet Exicor 150AT User Manual 71 ...
Страница 90: ...Appendix F Exicor 150AT Limited Warranty 80 Exicor 150AT User Manual ...