Chapter 4: Software Interface
40
Exicor 150AT User Manual
In order to access the auto calibration feature, check the
Enable Auto-Calibration
box. Auto-offset settings are defined in 2 ways:
11. By Time: take offsets on a time interval
12. By position: take offsets based on X-Y stage position
To change the number of single measurements averaged together for the auto-
calibration measurement, enter a new number in the field provided.
The most useful way to do offsets on the Exicor 150AT is by time. The position
method is more useful for other Exicor models. Both methods are described below.
Offsets by Time
When offsets are measured on a time basis, the system will take a measurement
offset before each scan then periodically during the scan. To use auto-calibration on
a time basis, select
Time
from the drop-down menu.
To change the time between offset measurements (during the scan), change the
value in the field provided (units: seconds).
To change the location for the automatic offset measurement, select
Change
Location
, edit the values for the x,y coordinate of the offset measurement.
Note:
The auto-calibration location must be a position on the sample
stage where no sample is present and the beam is unobstructed
(not blocked). Once set, click
OK
to save settings.
With the default settings as shown in the figure above, the system will make 5 offset
measurements and average them together for the air offset. It will do this every 20
minutes (1200 seconds) during each scan.
Содержание EXICOR 150AT
Страница 1: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F ...
Страница 2: ......
Страница 3: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F Software Version 6 5 ...
Страница 7: ...Table Of Contents Exicor 150AT User Manual v ...
Страница 16: ...Operator s Safety Summary Exicor 150AT User Manual 6 Figure S 1 Laser Beam Path ...
Страница 17: ...Laser Safety Exicor 150AT User Manual 7 Figure S 2 Compliance Labeling ...
Страница 21: ...Exicor 150AT User Manual 11 ...
Страница 26: ...Chapter 1 Overview of the Exicor Birefringence Measurement System 16 Exicor 150AT User Manual ...
Страница 29: ...Loading a Sample Exicor 150AT User Manual 19 ...
Страница 49: ...Configuring the System Exicor 150AT User Manual 39 Figure 4 6 Auto Calibration Settings Dialog Box ...
Страница 74: ...Chapter 4 Operating the System 64 Exicor 150AT User Manual ...
Страница 81: ...Cleaning the Cabinet Exicor 150AT User Manual 71 ...
Страница 90: ...Appendix F Exicor 150AT Limited Warranty 80 Exicor 150AT User Manual ...