Commissioning
84/146
Ø
Click
Yes
.
Ø
Wait until reset operation is finished and result is shown:
Ê
The message
Master configured successfully
appears.
Ê
The set IO-Link Wireless Master settings are used now.
Scan
Ø
Select
IO-Link Wireless
Master
in the left column of the netFIELD
Wireless Web Server.
Ø
Open the
Scan
tab.
Figure 39: Scan tab
Ø
Enter
TXPower
as decimal value: 31 (= maximum transmission power
of the device)
Ø
Click
Scan start
.
Ê
The scan result is displayed. The connected device is found.
Ê
The following scan result values are displayed:
Parameter
Scan result (example)
Note
Index
0
Unique ID
0x03,0xf3,0x00,0x00,0x01,0x72,0xc0,0x45,0x
cf
Copy/note the unique ID. This value is required for
port configuration.
Slot Type
Double slot
Revision ID
0x11
Port
“Select port”
Note: For a device featuring “Double slot” an even
port must be assigned.
Pairing
Pair (green)
Table 45: Scan result
Figure 40: Scan tab with result
NFD-3090-PNS-IOLM\W | Hardware, installation and commissioning
DOC201101UM03EN | Revision 3 | English | 2022-02 | Released | Public
© Hilscher 2020 - 2022