Table 1 Electromagnetic Emissions
Emission tests
Conformity
EMC environment - Guide
RF emission following
CISPR 11
Group 1
The test unit generates RF energy
only for internal use. Radiation
thus is low and it seems unlikely
that adjacent medical apparatus is
perturbed
RF emission following
CISPR 11
Class B
HIDREX tap water iontophoresis
devices are suitable for use in all
establishments, including domestic
establishments and those directly
connected to a public low-voltage
power supply network that sup-
plies buildings used for domestic
purposes.
Mains harmonics follow-
ing IEC61000-3-2
Class A
Emission of voltage
dips/ flicker following
IEC61000-3-3
Compliant
Table 2 Electromagnetic Immunity
Susceptibility
IEC 60601-test level
Actual level
ESD IEC 61000-4-2
+/-6kV cd
+/-8kV ad
+/-6kV cd
+/-8kV ad
Bursts
IEC 61000-4-4
+/-2kV mains
+/-1kV I/O
+/-2kV mains
+/-1kV I/O
Surges
IEC 61000-4-5
+/-1kV dm
+/-2kV cm
+/-1kV dm
+/-2kV cm
Voltage drops etc
IEC 61000-4-11
Reduction to
Reduction to
5 % for 10 ms / positive
amplitude
5 % for 10 ms / positive
amplitude
5 % for 10 ms / negative
amplitude
5 % for 10 ms / negativee
amplitude
40 % for 100 ms
40 % for 100 ms
30 % for 500 ms
30 % for 500 ms
0 % for 5000 ms
0 % for 5000 ms
H- field at 50/ 60 Hz
IEC 61000-4-8
3 A/m
3 A/m
Table 4 Electromagnetic Immunity - Life Support Equipment
Susceptibility
IEC 60601-test level
Actual level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3V
eff
150 kHz to 80 MHz
3V
eff
80 MHz to 2.5 GHz
3 V
3 V/m