DP450 Tap Water Iontophoresis System User Manual
Rev. 1.4 / 02-16-2017
14
Button
Pulse:
When PULSE is active, pulsed
current is used instead of direct
current.
LED ON = pulsed current
LED OFF = direct current
Button
ENTER key
Is used to go directly into Program-
ming Mode without waiting for 6
seconds.
Important:
The Setup Mode is accessible only directly after switching
on the device. If you leave the Setup Mode, you must turn
off the device briefly and then turn it on again in order to
access Setup Mode and alter the above treatment options.
In the factory, your DP450 is pre-set as follows:
Below are some recommendations for
initial
treatment, although most
users increase the voltage as they get more comfortable with the de-
vice. Each user should experiment to see what works for them and
how the treatment feels. It is recommended to increase the voltage as
it can be tolerated.
Pulsed Current: OFF
Time:
15 minutes
Dose:
4 Volts
Treatment area
Treatment parameter
HANDS
Pulsed Current ON
Time:
15 minutes
Dose:
20 Volts
FEET
Pulsed Current:
OFF
Time:
15 minutes
Dose:
30 Volts
Axillary (underarms)
Pulsed Current :
ON
Time:
15 minutes
Dose:
8 Volts