DP450 Tap Water Iontophoresis System User Manual
Rev. 1.4 / 02-16-2017
16
Hint:
The system saves these settings if Profile is active even if you
adjust parameters during a treatment. The device will automatical-
ly suggest these parameters
1
when a new treatment session is
started.
2) Starting the Treatment:
When treating hands or feet, place each hand or foot onto one towel
➷
each that
are covering the electrodes in the carrying case trays
➘
or ergonomic treatment
trays
➱
that are sufficiently filled with tap water according to the installation in-
structions noted previously. Make sure that your skin does not touch the elec-
trodes directly.
Do not forget to remove any jewelry!
Avoid direct contact between skin and the metal electrode!
To conduct an axillary treatment, place the pre-soaked axillary pads
✃
into your
underarms. Make sure the pads fit well and apply even pressure to maximize the
contact area.
When the built-in “immersion monitoring” sys-
tem recognizes that the circuit is closed, the
ACTIVE LED
lights up to indicate that the
treatment is active.
When treatment is active (ACTIVE
LED
light is ON) the actual dose
being currently administered is
displayed on the right of the dis-
play. Upon immersion or start of a
treatment, the dosage always
starts at zero and then slowly in-
creases to the desired dose previ-
ously set.
Note:
The dose measurement can be in milliamps or volts, depending on
your selection in the Setup Mode; it can also be toggled during an active
treatment, but NOT during Programming Mode.
Important:
If you experience any localized pain during a treatment, you
need to interrupt the session and cover the painful area with
petroleum jelly. Pressing any button will stop the automatic in-
crease of the treatment dose.
3) Adjusting treatment parameters:
At any time throughout the treatment session, you can change treatment
time and dose settings by pressing the respective push buttons.
If you interrupt the treatment by removing the treatment area from the water
source, the ACTIVE LED
goes off and the treatment timer is paused.
The