DP450 Tap Water Iontophoresis System User Manual      

Rev. 1.4 / 02-16-2017

 

16 

Hint: 

The  system  saves  these  settings  if  Profile  is  active  even  if  you 
adjust parameters during a treatment. The device will automatical-
ly  suggest  these  parameters

1

  when  a  new  treatment  session  is 

started. 

2) Starting the Treatment: 

When treating hands or feet, place each hand or foot onto one towel

 

each that 

are covering the electrodes in the carrying case trays

 or ergonomic treatment 

trays

 

that  are  sufficiently  filled  with  tap  water  according  to  the  installation  in-

structions  noted  previously.  Make  sure  that  your  skin  does  not  touch  the  elec-
trodes directly.  

Do not forget to remove any jewelry! 

Avoid direct contact between skin and the metal electrode!

 

To conduct an axillary treatment, place the pre-soaked axillary pads

 

into your 

underarms. Make sure the pads fit well and apply even pressure to maximize the 
contact area. 

When  the  built-in  “immersion  monitoring”  sys-
tem  recognizes  that  the  circuit  is  closed,  the 
ACTIVE  LED 

  lights  up  to  indicate  that  the 

treatment is active. 

When treatment is active (ACTIVE 
LED 

 light is ON) the actual dose 

being  currently  administered  is 
displayed  on  the  right  of  the  dis-
play. Upon immersion or start of a 
treatment,  the  dosage  always 
starts  at  zero  and  then  slowly  in-
creases to the desired dose previ-
ously set. 

Note:

      The  dose  measurement  can  be  in  milliamps  or  volts,  depending  on 

your  selection  in  the  Setup  Mode;  it  can  also  be  toggled  during  an  active 
treatment, but NOT during Programming Mode.  

Important:

  

If  you  experience  any  localized  pain  during  a  treatment,  you 
need to  interrupt  the session and cover the painful area with 
petroleum jelly. Pressing any button will stop the automatic in-
crease of the treatment dose.

 

3) Adjusting treatment parameters: 

At  any  time  throughout  the  treatment  session,  you  can  change  treatment 
time and dose settings by pressing the respective push buttons. 

If you interrupt the treatment by removing the treatment area from the water 
source, the ACTIVE LED 

 goes off and the treatment timer is paused.

 

The 

Содержание DP450

Страница 1: ...DP450 Tap Water Iontophoresis TWI System User Manual Included and optional accessories not shown...

Страница 2: ...llation Treatment Setup 11 SetupforHandsandFeetTreatment 11 SetupforAxillary Armpits Treatment 12 Conducting Treatments 12 PrimaryParameterSetup SetupMode 13 BeginTreatment ProgrammingMode 15 Importan...

Страница 3: ...therwise come in contact with the water source trays or pads Keeping such accessories on would lead to localized minor elec trical burns that are secondary to increased current densities Avoid applica...

Страница 4: ...est brain or sinuses Patients with suspected or diagnosed heart problems or epilepsy Side Effects Possible side effects or results of using this device include Mild dysesthesia tingling or burning Sho...

Страница 5: ...laundry instructions on the tag Intended Use Mechanism of Action Intended Use This tap water iontophoresis device is intended to treat hyperhidrosis 1 excessive sweating affecting hands feet and under...

Страница 6: ...that is attached to the black positive cable It is recommended to conduct all treatments in the initial treatment phase Phase 1 without changing polarity until one side reaches an acceptable dryness...

Страница 7: ...ient method of tap water iontophoresis treatment available Due to user comfort preferences this device gives the user the option to choose PC as an alternative Pulsed Current PC In PC the current alte...

Страница 8: ...s explained in the following sections Standard Accessories Optional Accessories Ergonomic treatment trays Axillary Treatment Kit 1 1 Axillary Treatment Kit includes a pair of small electrodes AX elect...

Страница 9: ...Iontophoresis System User Manual 9 Control UnitSide Rear Panel Elements 1 Control unit main ON OFF switch main power switch 2 Jacks for connecting the cable set to the treatment electrodes 4 Connector...

Страница 10: ...type of current pulsed or direct Programming Mode Decrease time Programming Mode Increase time Setup Mode Select Profile Indicates Profile preset has been selected Indicates Pulsed Current has been se...

Страница 11: ...sure to forcefully push the connectors all the way onto the necks of the treatment electrodes 5 Place one treatment electrode in each hard shell case1 or in each ergonomic treatment tray respectively...

Страница 12: ...to re soak the axillary pads Conducting Treatments For your comfort please observe the following two instructions before you begin a treatment session 1 BEFORE you start a treatment session and close...

Страница 13: ...ing Setup Mode and dur ing treatment also Important If no button is pressed within 6 seconds the device will au tomatically switch from Setup Mode to Programming Mode Hint When the device automaticall...

Страница 14: ...in in order to access Setup Mode and alter the above treatment options In the factory your DP450 is pre set as follows Below are some recommendations for initial treatment although most users increase...

Страница 15: ...eatment session you need to check and or set the time and dose treatment parameters 1 Setting treatment parameters If Profile has been selected the device uses the saved treatment parame ters 1 If the...

Страница 16: ...rsion monitoring sys tem recognizes that the circuit is closed the ACTIVE LED lights up to indicate that the treatment is active When treatment is active ACTIVE LED light is ON the actual dose being c...

Страница 17: ...rotection OTP Feature If your skin is dryer than normal 1 the dose cannot be increased protection against over treatment Your HIDREX system automatically checks your specific skin conductance If your...

Страница 18: ...display will show StOP and all LEDs will blink Please turn it off unplug the power supply from the wall and send the device with all electrical accessories back to us if this occurs Please refer to s...

Страница 19: ...evice we recommend the following steps after each use Care after each treatment Wash out towels or axillary pads thoroughly with clean water without addi tion of detergents or cleaning agents Squeeze...

Страница 20: ...ual check of medical device and accessories Protective conductor test in accordance with EN 60601 1 2007 Leakage current test in accordance with EN 60601 1 2007 Operational check of the medical device...

Страница 21: ...rough the following checklist and perform the operational check before you send the device for repair This can save you considerable cost and inconvenience Thank you Error Checklist Please go through...

Страница 22: ...erial for shipping Make sure the device is protected against impact inside the case and that packaging is suitable for shipping Prior to shipping do not forget to clean and dry the system and the acce...

Страница 23: ...s is reusable and not contaminated at end of life complies with WEEE Directive 2012 19 EU Manufacturer Hidrex GmbH Otto Hahn Str 12 Heiligenhaus Germany 42579 info hidrex de www hidrex de CE Mark ID o...

Страница 24: ...close to them 3 If unexpected events occurs remove treated area from the water and turn the powered device OFF 4 Be aware that adding accessories or components or modifying the powered device may make...

Страница 25: ...61000 3 2 Class A Emission of voltage dips flicker following IEC61000 3 3 Compliant Table 2 Electromagnetic Immunity Susceptibility IEC 60601 test level Actual level ESD IEC 61000 4 2 6kV cd 8kV ad 6k...

Страница 26: ...parts in an environmentally acceptable way If the device can no longer be used please dispose of it properly The national ordinances are to be heeded regarding any other parts Appliances that are mark...

Страница 27: ...tment voltage 40 V max Treatment current 30 mA max into a 1 k load Effective treatment current 30 mA max 5 mA reserved for the device Max output power 225 mW Pulsed current output Treatment voltage 40...

Страница 28: ...al Water Surface 112 13 0 26 0 31 Hand Feet Ergonomic Tray individual Water Surface 90 88 0 33 0 38 Hand Feet individual Electrode 122 13 0 24 0 28 Hand Feet individual Cover Towel 98 8 0 30 0 35 Axil...

Отзывы: