12
21/03/2019 Rev.2 (04/12/2019)
Step 8: Connecting the backrest and the legrest (on 3-motors beds only)
Unroll the electrical cable, connect the actuator to the main power supply and lift the bed to its maximum height.
Remove the cover of the power control box with a screwdriver or a standard key car for example.
Plug the actuators respecting the colours: blue in blue and orange in orange.
Insert the cables in their cable clamps and place back the power box cover.
On 3-motors bed, you need to connect the actuator for the legrest, on 2-motors bed, only the backrest actuator.
Step 9 :
Remote control
The Home Care bed is delivered with a precabled remote control.The electrical adjustments of the hi-low,
backrest and legrest (only for 3 motors version) can be done using the remote control.
• The remote control is equipped with a plastic hook to hang it.
• The spiral cable enables easy movement.
• The remote control is waterproof and washable.
• The remote control is equipped with a safety key to lock it
As long as you press the remote control button, the corresponding actuator will work. Adjustments are
possible in both directions, up or down.
The functions of the buttons are :
•
*
UP
•
*
DOWN
For safety reasons, there is a locking system built in the remote control
Special safety notes for use of side rails:
* Use only parts in perfect condition, do not use bent, cracked or damaged side rails.
* Use only original HERDEGEN SAS parts. Side rails from other suppliers won‘t fit correctly on the Home Care bed.
* Before using the side rails evaluate and take in account the physical condition of the patient :
• If the patient is very nervous, restless or confused don‘t use the siderails but other means such as safety belts for example.
• For very small patients, additional protective measures maybe necessary to reduce the distance between the tubes of the
siderail.
* Use mattresses adapted to the Home Care bed with a maximal height of 14 cm. Mattresses should be fire resistant according to DIN
597 Part 1 and 2 .
* Should specific antidecubitus mattresses be used, make sure that the side rail has an effective height of 22cm. If not, you must find
solutions in order to ensure the safety of the patient.
* Check regularly the various elements of the siderails (holders, pins, tubes..). In case of damages, immediately remove the side rail
and use a new one.
* Check every week the correct fixation of the side rails, screw again the knurled screws if necessary.
* Make sure, before using the side rail (folding or unfolding) that there are no risk of crushing or squeezing the patient‘s limbs. Thus,
you will avoid severe injuries.
If the upper rules are not followed, there are risks of severe injuries for the patient such as
crush , pinch point injuries and fall risks.
Beware: in case of use of incompatible mattresses, there are risks of severe injuries for the patient !
MATTRESS:
Please respect the medical prescription or recommendations regarding the size of the mattress. Check the user manual
if necessary.
Width of the bed's frame: 90cm
Suitable mattresses for the Ampli Plus medical bed:
Width: 86cm minimum - 90cm maximum Length: 196cm
High resiliency foam of 34Kg/m³ minimum
backrest
legrest
hi-low
Remote control
Clip to maintain cables
Cover
Backrest adjustment
Legrest adjustment
Hi-Low of the bed
Key of the TH2 remote control
(locking and unlocking):
Key inserted: remote control unlocked
Key removed: remote control locked
TH2
remote control
Power control box
Cables connection