29
21/03/2019 Rev.2 (04/12/2019)
In order to benefit from the guarantee and ensure the proper functioning of the bed, it is imperative to use only original spare parts
from HERDEGEN SAS.
For inquiries about spare parts and use of bed, please contact your authorized distributor or HERDEGEN SAS:
HERDEGEN SAS 5/7 Rue Freyssinet 77500 CHELLES FRANCE
Tel : +33 2.48.23.84.54
3.2 Cleaning and desinfection
3.2.1. General information
Follow the rules before starting the disinfection :
*Unplug the power cable from the wall outlet and protect it so that it doesn‘t come in contact with too much detergents.
*Ensure that all cables are correctly pluged into the central control box and close that box.
*Ensure that all electrical components are in perfect condition, so that detergents can‘t infiltrate the components.
*Dry the bed after cleaning and make sure before operating that no moisture remains on the electrical components.
*Do not use Kärcher or high pressure to clean the electrical components.
*If water or detergent has infiltrated some parts, take the bed out of service.
*
The non-compliance with the above rules and procedures may lead to injuries and damages on the
electrical components.
Start the cleaning :
*Remove the bed sheets.
*Clean all surfaces with a mild and environmentally safe cleaning solution. This also applies to the remote control.
*If the bed has visible contamination, it must be disinfected by wiping after cleaning. Use approved disinfectants for each surface
according to the products listed by the DGHM (German Organization for Hygiene and Microbiology).
*Disinfection of castors is only necessary in case of a visible contamination with potientally infectious material (blood, pus..)
Instructing the staff
Qualified staff must be taught how to clean and disinfect the Home Care bed. The following points must be respected :
*The Home Care bed musn‘t be contaminated during its transport to the patient‘s home.
*We advise to clean and disinfect the bed immediately after disassembling. We recommend using only approved disinfectant in the
approved concentration.
*The cleaning staff must be equipped with handgloves and clothes.
*The cleaning staff must use clean towels to clean and disinfect, those must be immediately laundered after use.
*Once the disinfection is finished, the cleaning staff must wash and disinfect their hands before going to the next patient. The staff
should be equipped with an appropriate disinfectant. The immediate cleaning-disinfection of the bed on place avoid transporting
dirty beds and potential infectious spores.
*Home Care beds must be covered when stored to avoid external and accidental pollution.
3.2.2. Cleaning and desinfection
Follow these instructions to keep the Home Care bed in a good state for a longer time.
*Surfaces must be in perfect condition, scratches must be immediately repaired to avoid the penetration of moisture.
*We advise to use a moist cloth with a mild commercial household cleaner.
*Use only listed disinfectant to wash down the bed, please refer to German Organization for Hygiene and Microbiology.
*Do not use abrasive agents or scouring agents to clean the bed, you could damage the surface.
*Mix the disinfectant with cold water to avoid vapors.
*Wear gloves to avoid direct skin contacts.
*Ventilate the room after disinfection.
*Don‘t use spray to clean the Home Care bed as you could inhalate it.
*Don‘t use alcohol to clean larger areas.
Distributør i Norge:
Gymo AS - www.gymo.no
(+47) 4000 3522 - [email protected]