Device Settings
• The MyoVista Device recognizes only the first supported partition on a USB storage media . Do
not use storage media with multiple formatted partitions .
• Store backed up databases in compliance with HIPAA controls and policies for handling
sensitive patient information .
• Deleting a database deletes all patient information and reports permanently, but does not
affect system and device settings .
3 . Select Backup icon . The Backup Database screen appears, as shown in Figure 5-33 .
Note: If multiple USB storage media are connected, select one from the drop-down menu. For
information about USB mapping, see “Rear Panel” on page 26.
Figure 5-33. Backup Database Screen
4 . Within the “Notes” field, add an optional note .
Tip: We recommend you add a note in the remark box to help identify the database backup
correctly.
5 . Select the Backup icon to back up the existing database or select Cancel icon to not back up the
database . If you select Backup icon, the following message appears:
Backup Successful
6 . Select the OK icon to display the Database Information screen or select Cancel icon to return to
the Database Information screen .