USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Device Settings
User Management
User Management allows administrative users to enable and configure user authentication and user
accounts for the device .
Any user is able to change their own user details from this screen, including display name, username,
password, and hint .
Add User
1 . Under User Settings, select the User Management icon . The Login Configuration screen appears
as shown in Figure 5-52 .
Figure 5-52. Login Configuration Screen
Note: By default, the device ships with User Authentication disabled.
2 . Select the User Authentication button to enable authentication .
Note: Once authentication is enabled, make sure to add at least one administrator account.
Note: If an admin account cannot be accessed (
forgot password, “admin” account not
configured, etc.)
the administrator features of the device can be temporarily accessed by
contacting technical support at HeartSciences, where a temporary admin password can be
provided.